Source container FQG61, FQG62
Hauser
43
8. Only for source containers with order code 670 "Additional function", option WA "Density
measurement > Fixation ON": screw in setscrew using Allen key. The current switching state is
indicated by the arrow (B) ("EIN-ON" or "AUS-OFF").
5 mm
B
7.
8.
A0033940
9. Put the cover back on. The "EIN-ON" position must be visible.
9.
A0033941
Switching OFF the radiation
Proceed in the same way to switch the radiation off. To switch the radiation off, turn the insert 180°
in the clockwise direction.
Reading the switch state:
• Radiation ON: the "EIN - ON" sign is visible. The arrow points to "EIN - ON"
• Radiation OFF: the "AUS - OFF" sign is visible. The arrow points to "AUS - OFF"
FQG61/FQG62; order code
020, option B
Switching ON the radiation
L
CAUTION
Safety instructions for switching ON the radiation
‣
Before switching on the radiation, make sure that no-one is in the radiation zone (or inside the
product vessel).
‣
The radiation may only be switched ON by properly instructed staff.
Switching ON the radiation
L
WARNING
Do not press the sealed locking pin (B). Rotating the insert over the locking pin leads to the
removal position of the source insert.
‣
Do not remove the safety bracket (A).
1. Remove the padlock.
2. Release the locking screw (optional).