Flowfit CYA27
Installation
Hauser
25
5.5
Installing sensor in assembly
5.5.1
Disinfection sensor
Please note the following during installation:
‣
Configure the minimum flow rate.
‣
If the medium is fed back into an overflow basin, pipe or similar, the resulting
counterpressure on the sensor must not exceed 2 bar (29 psi) bar absolute and must
remain constant.
‣
Avoid negative pressure at the sensor, e.g. due to medium being returned to the suction
side of a pump.
‣
To avoid buildup, heavily contaminated water should also be filtered.
1
1.
2.
3.
4.
A0043536
1. The assembly is supplied to the customer with a union nut screwed onto the
assembly: unscrew union nut from assembly.
2. The assembly is supplied to the customer with a dummy plug inserted in the
assembly: remove dummy plug and O-ring (1) from the assembly.
3. Slide the sensor with the adapter for Flowfit CYA27 into the opening in the assembly.
4. Screw union nut onto assembly on block.