Installation
Flowfit CYA27
16
Hauser
Number of modules
1
2
3
Spacing between drill
holes
mm (in)
120 (4.73)
180 (7.09)
240 (9.45)
The mounting materials required to secure the device to the wall are not supplied.
1. Provide the mounting materials to secure the device to the wall (screws, wall plugs)
onsite.
2. Use mounting material that is suited to the wall substrate.
M6
240 (9.45)
A0043508
4
Engineering unit: mm (in)
5.2.3
Mounting the assembly with wall holder unit
With the wall bracket, it is possible to select up to six modules. It is then possible to
remove individual modules while the rest of the assembly remains securely in the bracket.
Different drill holes allow, for example, the use of the Flowfit CCA250 hole pattern.
Optional accessories, comprising a wall bracket with securing clips for assemblies with
1 to 6 modules.
70 (2.76)
80 (3.15)
70 (2.76)
240 (9.45)
320 (12.6)
153.6 (6.05)
Ø9
(4
x)
7
(0
.2
8)
Ø9
(4
x)
80 (3.15)
A0043220
5
Dimensions of wall bracket. Engineering unit: mm (in)
The drill holes highlighted in color correspond to those of assembly CCA250, which
can be reused.