CE 4
Maintenance and cleaning
Hauser
19
If an automatic rinse system is available and the pump is switched on, the complete
pipework system including sensors can be rinsed in presettable time cycles using water
from the local supply. Rinse cycle and time for hose and sensors must be matched to
suit individual application and can be individually set up. Individual set-up points are
available using numbered "Block numbers" (e.g. B01=Sensor rinse interval set up
position). For controller set-up
→
Chap. 7.3.4
Suction hose rinse:
Fig. 11:
Time cycles for automatic suction hose rinse
Sensor rinse:
Fig. 12:
Time cycles for automatic sensor rinse
on
off
A
B / C
I
K / H
B
C
B
H
on
off
D
E
I
K
Pos.
Function
Set up address
Time interval
Factory default
A
Start hose rinse
B01
(0...99h)
24h
B
Hose rinse time
B02
(0...99s)
45s
C
pump rinse time
B03
(0...99s)
30s
D
Start sensor rinse
B04
(0...99h)
18h
E
Sensor rinse time
B05
(0...99s)
60s
H
Hold time after suction hose rinse
B08
(0...99s)
20s
I
Pump status
K
Hold (transmitter)