Installation and commissioning
HS 55 (Batch Mode)
28
Figure 17Hose routing for the "Hg with enrichment" mode
Selecting the cell
}
"Hydride"
mode: use the quartz cell.
}
Insert the cell into the cell unit of the AAS device and close it with the quartz win-
dows.
}
"Hg with/without enrichment"
modes: optionally select the tulip-shaped Hg cell. Al-
ternatively, the quartz cell can be used.
}
Insert the cell into the cell unit and close it with the springs.
4.2
Retrofitting the Hg Plus Upgrade Modul
WARNING
Risk of electric shock
There is a risk of touching live components and suffering an electric shock when con-
verting the Hg/hydride system.
¡
Before the conversion: switch off the Hg/hydride system and the AAS at the power
switch.
¡
Disconnect the power plug from the socket.
¡
Disconnect all connection cables to the AAS and the cell unit.
Software tool for the conver-
sion
}
Insert the CD included in the delivery into the PC. Start the HS Wizard program and
follow the on-screen instructions.
}
Select the spectrometer used.
}
Select the starting configuration of the Hg/hydride system.
}
Select the target configuration of the Hg/hydride system with Hg enrichment.