Installation and commissioning
HS 55 (Batch Mode)
24
Figure 12Hg/hydride system installed on the AAS
}
Place the Hg/hydride system to the right of the AAS device or onto a table next to
the AAS device.
Electrical connection and inter-
faces
}
Connect the cell unit electrically to the Hg/hydride system:
– Connect the heating cable to the "cell heating" connection.
– Connect the temperature sensor cable to the "cell sensor" connection.
– Fasten the grounding of the sensor cable on the connection panel of the Hg/hy-
dride system with the knurled head screw.
}
Connect the functional module to the AAS: the AAS device provides the voltages
(+5 V/+24 V) for the functional module.
– Connect the "AAS" plug of the twin cable to the AS or Sampler Flame socket on the
AAS device.
– Connect the "HS" D-sub socket of the thinner cable of the twin cable to the "input
5 V/24 V" connection of the Hg/hydride system.
– The round green "AS" socket of the twin cable is not used.
– Connect the signal cable to the HS or the Hydridsystem connector of the AAS de-
vice. Connect the other end of the cable to the "AAS/RS 232" interface of the Hg/
hydride system.