CyBio Carry
Maintenance and care
41
7
Maintenance and care
The operator may not undertake any service or maintenance work to this device and
its components other than that specified in these instructions.
Observe the information in the "Safety instructions" section for all maintenance work.
Compliance with the safety instructions is a prerequisite for the error-free operation
of the device. Always observe all warnings and instructions that are displayed on the
device itself or indicated by the control software.
To ensure faultless and safe functioning, Analytik Jena recommends an annual in-
spection and servicing by its Service department.
7.1
Maintenance overview
Maintenance interval
Maintenance task
Once per week and
when contaminated
Clean the device housing.
Every six months
¡
Check the electrical components and cables for damage.
¡
Protective conductor test (only by a qualified electrician)
7.2
Cleaning the device
Use a soft cloth dipped in mild soapy or disinfectant solution to clean the device
housing.
Never use cleaning powder, paint thinners or solvents like petrol or acetone to clean
the device! These can corrode the housing surface.
Use a lint-free cloth with a cleaning agent / disinfectant recommended by WHO
guidelines and not prohibited by this manual (e.g., Incidin Liquid from ECOLAB) to
clean the CyBio QuadStack and all accessories only cleanable via wipe disinfection.
Chemical resistance
Spraying the device with disinfectant spray or similar can be dangerous and is pro-
hibited for this reason. Sprays contain gases which may ignite.
Contamination and natural wear of assemblies leads to higher stress on the device
and thus to a higher probability of device failure. Check for signs of wear on assem-
blies under mechanical strain and have these replaced when necessary.