![endaq Slam Stick C User Manual Download Page 27](http://html1.mh-extra.com/html/endaq/slam-stick-c/slam-stick-c_user-manual_2406599027.webp)
User Manual
Slam Stick User Manual
Version No. 2.0
27
Figure 10: The Triggers tab lets the user set a variety of triggers for starting a recording.
In order to set each trigger value, the box next to the trigger must be checked. All triggers can be
used simultaneously with other triggers, and in applications where multiple triggers are set, Slam
Stick will wait for each trigger setting to be accomplished before a recording begins. But some
triggers have priority, in order they appear on the screen. For example if a delay is programmed
along with a temperature trigger, and an acceleration trigger, the device will first wait the delay
period, then start looking for the temperature trigger condition to be met. Only after the
temperature trigger has been met will the device then begin waiting for the acceleration trigger.