background image

Users’ Manual 

EAN-90M Tiltmeter 

 

 

Page

 | 

6

 

3  MEASUREMENT OF TILT 

The output of model EAN-90M tilt meter can be read by Encardio-rite model EDI-53UTM portable digital 
read-out unit/datalogger. The EAN-90M can also be read or logged from a remote location by automatic 
data acquisition system like Encardio-rite model EDAS-10. The tilt meter is supplied with 2 m built-in six 
core cable. The cable can be terminated or extended to the nearest measurement station through a suitable 
junction box.  

Note: 

  For automatic logging by Encardio-rite model ESDL-30 datalogger, model EAN-92M Tiltmeter with 

in-built SDI-12 digital interface is available.  

When using with Encardio-rite model EDI-53UTM portable digital read-out unit, a break-out box (special 
junction box) having suitable circular connector is available from Encardio-rite.  

3.1  Wiring details 
 

Cable colour   Description 

 

Red 

+ 12V DC 

 

Black 

- 12V DC 

 

Green 

0 V (Power supply common) 

 

Blue 

Output Channel A 

 

Brown 

Output Channel B  
(only for biaxial tilt sensors) 

 

White 

Output Common 

3.2  Sign convention 

Careful orient tilt sensor during installation. Make a note of the orientation. Uni-axial sensor measures tilt 
only along axis ‘A”. Figure 4 shows a view from top and also convention used for direction/output signal 
polarity.  

3.3  Connecting sensor to Encardio-rite Digital readout unit 

Tilt meter can be connected to Encardio-rite model EDI-53UTM 
portable digital read-out unit/datalogger through a break-out 
box.  Break-out box provides the mating circular connector to 
read-out unit. The cable can also be extended through the 
break-out box. The breakout box is also equipped with lightning 
arrestor.  

Encardio-rite model EDI-53UTM portable digital read-out 
unit/datalogger reads sine of tilt angle when parameter of typical 
tilt meter is fed. To read sine of tilt angle, set-up        EDI-53UTM 
channel as follows: 

Unit [UNITS]   

No units (key 9) 

Initial reading [IR] 

Set to zero 

Gauge factor [GF] 

2/tilt meter gage factor in volts/sin 90° 

Coeff of X²   

Set to zero 

Decimal point [DP] 

As required (recommended value is 4) 

Tilt angle can be calculated by sin

-1

(observed value). 

For bi-axial tilt meter, set-up two channels (one for A-axis and one for B-axis) in similar manner. 

Figure 4: Tiltmeter sign  

B

+

-

A

+

-

Tilt

Tilt

Breakout box 

Summary of Contents for EAN-90M

Page 1: ...gh ingenuity USERS MANUAL ENCARDIO RITE ELECTRONICS PVT LTD A 7 Industrial Estate Talkatora Road Lucknow UP 226011 India P 91 522 2661039 42 Email geotech encardio com www encardio com International U...

Page 2: ...f operation 2 2 2 Tool and accessories 2 2 3 Installation of Tiltmeter 2 2 3 1 Installation on vertical surface 3 2 3 2 Installation on horizontal surface 4 2 3 3 Installation on roof 5 3 MEASUREMENT...

Page 3: ...landslide areas Monitoring tunnels for convergence and other movement Monitoring safety of structures around zones of excavation or tunneling Monitoring deflection in bridges and struts under differe...

Page 4: ...a low thermal sensitivity The tilt meter can be fixed to any vertical surface horizontal floor or ceiling by means of suitable mounting accessories consisting of brackets and anchors These are availa...

Page 5: ...on the anchor and tighten slightly 4 Fix tilt meter on the mounting bracket with two M6 x 20 mm Allen bolts 5 Connect tilt meter with read out unit 6 Tighten anchor and Allen bolts nuts so that the t...

Page 6: ...nt 4 Insert anchors in drilled holes 5 Mount tilt meter bracket on the anchors and tighten slightly 6 Fix the tilt meter on mounting bracket with two M6 x 20 mm Allen bolts 7 Connect tilt meter with r...

Page 7: ...on anchor and tight it slightly 4 Fix the tilt meter on the mounting bracket with two M6 x 20 mm Allen bolts 5 Connect tilt meter with read out unit 6 Tighten anchor and Allen bolts nuts so that the t...

Page 8: ...3 2 Sign convention Careful orient tilt sensor during installation Make a note of the orientation Uni axial sensor measures tilt only along axis A Figure 4 shows a view from top and also convention us...

Page 9: ...ator is therefore required as shown in diagram below Use EDAS 10 function P2 to read output of tilt sensor in terms of 0 5 volts The multiplier and offset should be kept as 1 and 0 respectively To get...

Page 10: ...8367 0 0008 0 019 0 0 0000 0 0000 0 0000 0 0005 0 0000 0 0000 0 000 3 0 0524 0 8376 0 8375 0 8371 0 8376 0 0001 0 002 6 0 1046 1 6728 1 6726 1 6725 1 6730 0 0004 0 010 9 0 1565 2 5032 2 5029 2 5028 2...

Reviews: