3.5 Sample Test Certificate
TEST
CERTIFICATE
Item
:
Tilt meter (Uniaxial)
Date : 14.03.2018
Model
:
EAN-90M
Temperature : 27 ºC
Range
:
±15º
Serial no.
:
xxxxxxx
Next calibration due on :13.03.2019
Test data
Test
SinA
*Calculated
**Ideal
Observed
Offset
Error
Non-
position
output
output
output
corrected output
'A' axis
conformance
Arc degrees
'A' axis
(V
1
)
(V
2
)
(V
3
)
(V
4
)
( V
4
- V
2
)
% fs
(A)
Volts
Volts
Volts
Volts
Volts
15
0.2588
4.1408
4.1403
4.1404
4.1399
0.0004
0.010
12
0.2079
3.3264
3.3260
3.3259
3.3254
0.0006
0.014
9
0.1565
2.5032
2.5029
2.5027
2.5022
0.0007
0.017
6
0.1046
1.6728
1.6726
1.6721
1.6716
0.0010
0.024
3
0.0524
0.8376
0.8375
0.8372
0.8367
0.0008
0.019
0
0.0000
0.0000
0.0000
0.0005
0.0000
0.0000
0.000
-3
-0.0524
-0.8376
-0.8375
-0.8371
-0.8376
0.0001
0.002
-6
-0.1046
-1.6728
-1.6726
-1.6725
-1.6730
0.0004
0.010
-9
-0.1565
-2.5032
-2.5029
-2.5028
-2.5033
0.0004
0.010
-12
-0.2079
-3.3264
-3.3260
-3.3260
-3.3265
0.0005
0.012
-15
-0.2588
-4.1408
-4.1403
-4.1402
-4.1407
0.0004
0.010
Max non-conformance (% fs) :
0.02
Sensor gauge factor :
15.998
Volts/Sin(90)
Calculation of tilt value (arc degree) :
SinA = Observed output / gauge factor
A = Sin
-1
(observed output / gauge factor)
Wiring colour code :
Wire colour
Signal
Red
+12 V (supply )
Black
- 12 V (supply )
Green
0 V (supply )
Blue
Output 'A' axis
Brown
N/A
White
Output common
Note : * Calculated output Voltage (V
1
) worked out based on nominal gauge factor of 16.000 V/g
(i.e. 16V X Sin A).
**
Ideal output Voltage (V
2
)' calculated from sine curve passing through sensitivity calibration points
(@ ±15º) .
Tested by