3.5  Sample Test Certificate 

TEST

 

CERTIFICATE

 

 

 

 

 

 

 

 

 

Item 

Tilt meter (Uniaxial) 

 

Date             :  14.03.2018 

Model  

EAN-90M 

 

 

Temperature  :  27 ºC 

Range 

±15º 

 

 

 

 

 

Serial no. 

xxxxxxx 

 

 

 

 

 

Next calibration due on :13.03.2019 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Test data

 

  

  

  

  

  

  

  

Test  

SinA 

*Calculated  

**Ideal 

Observed  

Offset  

Error 

Non- 

position 

 

 output 

 output 

output 

corrected output 

'A' axis 

conformance 

Arc degrees  

 

 

 

'A' axis 

 

 

 

 

 

(V

1

(V

2

(V

3

(V

4

( V

4

 - V

2

 ) 

% fs 

(A) 

  

Volts 

Volts 

Volts 

Volts 

Volts 

  

 

 

 

 

 

 

 

 

15 

0.2588 

4.1408 

4.1403 

4.1404 

4.1399 

0.0004 

0.010 

12 

0.2079 

3.3264 

3.3260 

3.3259 

3.3254 

0.0006 

0.014 

0.1565 

2.5032 

2.5029 

2.5027 

2.5022 

0.0007 

0.017 

0.1046 

1.6728 

1.6726 

1.6721 

1.6716 

0.0010 

0.024 

0.0524 

0.8376 

0.8375 

0.8372 

0.8367 

0.0008 

0.019 

0.0000 

0.0000 

0.0000 

0.0005 

0.0000 

0.0000 

0.000 

-3 

-0.0524 

-0.8376 

-0.8375 

-0.8371 

-0.8376 

0.0001 

0.002 

-6 

-0.1046 

-1.6728 

-1.6726 

-1.6725 

-1.6730 

0.0004 

0.010 

-9 

-0.1565 

-2.5032 

-2.5029 

-2.5028 

-2.5033 

0.0004 

0.010 

-12 

-0.2079 

-3.3264 

-3.3260 

-3.3260 

-3.3265 

0.0005 

0.012 

-15 

-0.2588 

-4.1408 

-4.1403 

-4.1402 

-4.1407 

0.0004 

0.010 

  

  

  

  

  

  

  

  

 

 

 

 

 

 

 

 

 

 

 

 

Max non-conformance (% fs)    : 

0.02 

 

 

 

 

 

 

 

 

Sensor gauge factor  : 

15.998 

Volts/Sin(90) 

 

 

 

 

 

 

 

 

 

 

 

Calculation of tilt value (arc degree)  :   

 

 

 

 

 

 

SinA =  Observed output / gauge factor 

 

 

 

 A =  Sin

-1

(observed output / gauge factor) 

 

Wiring colour code : 

 

 

 

 

 

 

 

 

Wire colour 

Signal 

 

 

 

 

 

Red 

+12 V (supply ) 

 

 

 

 

 

Black 

- 12 V (supply ) 

 

 

 

 

 

Green 

  0 V (supply ) 

 

 

 

 

 

Blue 

 Output 'A' axis 

 

 

 

 

 

Brown 

N/A 

 

 

 

 

 

White 

Output common  

 

 

 

Note :         *  Calculated output Voltage (V

1

) worked out based on nominal gauge factor of 16.000 V/g  

(i.e. 16V X Sin A). 

** 

Ideal output Voltage (V

2

)' calculated from sine curve passing through sensitivity calibration points  

(@ ±15º) . 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Tested by 

 

 

 

 

 

 

 

 
 

Summary of Contents for EAN-90M

Page 1: ...gh ingenuity USERS MANUAL ENCARDIO RITE ELECTRONICS PVT LTD A 7 Industrial Estate Talkatora Road Lucknow UP 226011 India P 91 522 2661039 42 Email geotech encardio com www encardio com International U...

Page 2: ...f operation 2 2 2 Tool and accessories 2 2 3 Installation of Tiltmeter 2 2 3 1 Installation on vertical surface 3 2 3 2 Installation on horizontal surface 4 2 3 3 Installation on roof 5 3 MEASUREMENT...

Page 3: ...landslide areas Monitoring tunnels for convergence and other movement Monitoring safety of structures around zones of excavation or tunneling Monitoring deflection in bridges and struts under differe...

Page 4: ...a low thermal sensitivity The tilt meter can be fixed to any vertical surface horizontal floor or ceiling by means of suitable mounting accessories consisting of brackets and anchors These are availa...

Page 5: ...on the anchor and tighten slightly 4 Fix tilt meter on the mounting bracket with two M6 x 20 mm Allen bolts 5 Connect tilt meter with read out unit 6 Tighten anchor and Allen bolts nuts so that the t...

Page 6: ...nt 4 Insert anchors in drilled holes 5 Mount tilt meter bracket on the anchors and tighten slightly 6 Fix the tilt meter on mounting bracket with two M6 x 20 mm Allen bolts 7 Connect tilt meter with r...

Page 7: ...on anchor and tight it slightly 4 Fix the tilt meter on the mounting bracket with two M6 x 20 mm Allen bolts 5 Connect tilt meter with read out unit 6 Tighten anchor and Allen bolts nuts so that the t...

Page 8: ...3 2 Sign convention Careful orient tilt sensor during installation Make a note of the orientation Uni axial sensor measures tilt only along axis A Figure 4 shows a view from top and also convention us...

Page 9: ...ator is therefore required as shown in diagram below Use EDAS 10 function P2 to read output of tilt sensor in terms of 0 5 volts The multiplier and offset should be kept as 1 and 0 respectively To get...

Page 10: ...8367 0 0008 0 019 0 0 0000 0 0000 0 0000 0 0005 0 0000 0 0000 0 000 3 0 0524 0 8376 0 8375 0 8371 0 8376 0 0001 0 002 6 0 1046 1 6728 1 6726 1 6725 1 6730 0 0004 0 010 9 0 1565 2 5032 2 5029 2 5028 2...