AMETEK CTS
BS 200N100
Operating manual
V 1.12
12 / 21
5.4.
Measuring results
5.4.1. Voltage drop in dependence of the test load
DUT current
Nominal voltage drop
voltage drop
( typical value )
1
A
≤ 2.0
0.01
V
5
A
≤ 2.0
0.04
V
10
A
≤ 2.0
0.07
V
25
A
≤ 2.0
0.17
V
50
A
not specified
0.35
V
100 A
≤ 1.0
EM Test specs
.
0.96
V
➔
Specification values (measured between Power IN+ und DUT+)
5.4.2. Measuring results with verification set-up as per Standard
All measuring values indicated as well as the curve illustrated have been recorded with a serial model.
Reference values measured with BS 200N100
Input Voltage:
13.5 V
Output Voltage:
13.30 V
Upeak:
-
1030 V
tf (90% - 10%):
327.4 ns
Figure 5.3. : Pulse parameter
5.4.3. Verification
Verification as per ISO 7637-2 Ed3 (2011-03) is defined at 13.5V only. Verification must be performed with
calibration load CABS: 0.6
in series with 50
H; R
shunt
not used
Verification with CABS
Va
output voltage
tf
90% - 10%
fall time
With standard load
+13.5V
10%
300ns
20%
Function
Setting
Switch Mode
NO / NC
Trigger Mode
Man.Ext / Auto
Manual
OFF / ON
Test
OFF / ON
Pulse Repetition
1s
Pulse Duration
10ms
Example for the switch off impulse
Figure 5.4.: Test load at 13.5V