Battery replacement
Before replacing the battery the measuring pins must be disconnected from of the circuit
or a device. If the low battery symbol is displayed on the display (
), it is necessary to
replace the battery as soon as possible.
1. Unscrew the screws with a suitable screwdriver on the backside of the multimeter
and remove the cover.
2. Remove the old battery and replace it with a new one with a specified dimension
(12 V, type 23 A).
3. Put the cover on and screw it back.
Safety fuse replacement
Before replacing the safety fuse the measuring pins must be disconnected from of the
circuit or device. The safety fuse blows usually if the prescribed measurement procedure
is not followed down.
1. Unscrew the screws with a suitable screwdriver on the backside of the multimeter
and remove the cover.
2. Remove the blown safety fuse and replace it with a new one with a specified dimension
(F 250 mA/300 V AC fuse, 5 × 20 mm).
3. Put the cover on and screw it back.
Note: For a constant overload protection use a safety fuse with a specified current and voltage.
Maintenance instructions of the multimeter
Warning
• Do not attempt to repair the multimeter or modify it in any way, if you are not qualified
to execute such an activity and if you do not have the appropriate calibration devices.
• To prevent electric shock, ensure that no liquid water enters the interior of the
multimeter!
• Disconnect the measuring pins from the circuit before opening the cover of the
multimeter.
• Clean the body of the multimeter regularly with a damp cloth and a mild detergent.
Cleaning should be done only when the multimeter is switched off and disconnected.
• Do not use solvents or abrasives for cleaning!
• If you are not using the multimeter for a long time, turn it off and remove the battery.
• Do not store the multimeter in a place where there is a high humidity and temperature,
or in the environment with a strong magnetic field!
• When replacing the safety fuse, make sure a new fuse is of the same type and the
same range as the original one. Safety fuse: (F 250 mA/300 V), type F, Ø 5 × 20 mm.
The purchased device is under a warranty concerning a material quality and multimeter
workmanship. This warranty is valid for two years. The warranty does not include a battery
and a safety fuse. A complaint will not be accepted if the defect of the device is caused by
its improper use or exposure to extreme conditions.
Information on the backside of the multimeter:
Warning
Read all instructions before using this device. To prevent a possible electric shock, disconnect
the measuring leads from the power supply before opening the multimeter.
This device is not intended for use by a person (including children) whom a reduced physical,
sense or mental inability or a lack of experience and knowledge inhibits from a safe using of