6.4.1
Signal limits
Table 6-1: Maintenance Channels' Limits
Channel
Short range
Medium range
Long range
17 ft.
(5 m)
52 ft.
(16 m)
46 ft.
(14 m)
131 ft.
(40 m)
197 ft.
(60 m)
Reference
2 V
gain 0
1.5 V
gain 2
2 V
gain 0
1 V
gain 1
1 V
gain 2
Signal 1 & 2
2 V
gain 0
1.5 V
gain 2
2 V
gain 0
1 V
gain 1
1 V
gain 2
Ratio 1 & 2
0.6 to 1.4
NQRat 1 & 2
0.97 to 1.03
ppm.m
0 ppm.m
Temperature
Up to 25 °C beyond ambient temperature
Voltage
32 Vdc > V > 18 Vdc
(1)
The minimum distance, as defined on the model number.
(2)
Half the maximum distance, as defined on the model number.
(3)
The maximum distance, as defined on the model number.