2.3.6
HART
®
protocol
The Rosemount 936 uses the HART protocol.
This is a field communications protocol used to communicate between intelligent field
instruments and the host system.
HART is the global standard for smart instrumentation, and most smart field devices
installed in plants worldwide are HART-enabled. HART technology is easy to use and very
reliable.
Through the HART connection, the detector can:
•
Set up
•
Troubleshoot
•
Display health and status
You can connect the HART communication on the 0-20 mA line or thought the intrinsically
safe (I.S.) connection with a standard Field Communicator loaded with Emerson's host
software and attached by a special harness.
2.3.7
Modbus
®
RS-485
For more advanced communications, the Rosemount 936 has an RS-485 Modbus-
compatible output that provides data communication.
This feature enables simple commissioning, maintenance, and configuration.
2.3.8
Tilt mount
The stainless steel tilt mount provides a small installation footprint that can conform to
limited space constraints, while the sturdy construction maintains alignment even in
constant vibration.
The X and Y axis adjustments provide quick and easy alignment for installation and
maintenance.
2.4
Product certifications
2.4.1
ATEX and IECEx
The Rosemount 936 is approved per:
Ex II 2(2) G D
Ex db eb ib [ib Gb] IIB+H
2
T4 Gb
Ex tb [ib Db] IIIC T135 °C Db
Ta = -55 °C to +65 °C
Technical description
Reference Manual
October 2021
00809-0100-4036
12
Emerson.com/Rosemount
Summary of Contents for Rosemount 936
Page 4: ...4 ...
Page 8: ...Product overview Reference Manual October 2021 00809 0100 4036 8 Emerson com Rosemount ...
Page 24: ...Operating modes Reference Manual October 2021 00809 0100 4036 24 Emerson com Rosemount ...
Page 30: ...Specifications Reference Manual October 2021 00809 0100 4036 30 Emerson com Rosemount ...
Page 50: ...Installation Reference Manual October 2021 00809 0100 4036 50 Emerson com Rosemount ...
Page 68: ...Wiring configurations Reference Manual October 2021 00809 0100 4036 68 Emerson com Rosemount ...
Page 72: ...Support Reference Manual October 2021 00809 0100 4036 72 Emerson com Rosemount ...
Page 74: ...Short range model Reference Manual October 2021 00809 0100 4036 74 Emerson com Rosemount ...