Quick Installation Guide
00825-0100-4750, Rev BB
January 2013
Rosemount 8750WA
3
S
TEP
1: P
RE
-I
NSTALLATION
Before installing the Rosemount 8750WA Magnetic Flowmeter Transmitter, there are
several pre-installation steps that should be completed to make the installation process
easier:
• Identify the options and configurations that apply to your application
• Set the hardware switches if necessary
• Consider mechanical, electrical, and environmental requirements
Mechanical Considerations
The mounting site for the Rosemount 8750WA transmitter should provide enough room for
secure mounting, easy access to conduit ports, full opening of the transmitter covers, and
easy readability of the LOI screen (see Figure 1 and Figure 2).
If the Rosemount 8750WA transmitter is mounted separately from the sensor, it is not
subject to limitations that might apply to the sensor.
NOTE:
*M20 and PG 13.5 connections are available with the use of threaded conduit adapters.
Figure 1. Rosemount 8750WA32E Dimensional Drawing
5.82
(148)
6.48 (165)
7.49 (190)
LOI Cover
4.97
(126)
8.81
(224)
3.00
(76)
3.07
(78)
4.97
(126)
1
/2”-14 NPT* (3 places)
1
/2”-14 NPT, (2 places)*
4750RevBBQIG.fm Page 3 Friday, January 11, 2013 6:26 PM