Quick Installation Guide
00825-0100-4750, Rev BB
January 2013
Rosemount 8750WA
14
S
TEP
5: G
ROUNDING
Use Table 5 to determine which process grounding option to follow for proper installation.
The sensor case should be earth grounded in accordance with national and local electrical
codes. Failure to do so may impair the protection provided by the equipment.
Table 5. Process Grounding Installation
Process Grounding Options
Type of Pipe
Grounding
Straps
Grounding
Rings
Grounding
Electrode
Lining
Protectors
Conductive
Unlined Pipe
See Figure 10
Not Required
Not Required
See Figure 11
Conductive Lined
Pipe
Insufficient
Grounding
See Figure 11
See Figure 10
See Figure 11
Non-Conductive
Pipe
Insufficient
Grounding
See Figure 12 on
page 15
See Figure 13 on
page 15
See Figure 12 on
page 15
Figure 10. Grounding Straps or Grounding Electrode in Lined Pipe
Figure 11. Grounding with Grounding Rings or Lining Protectors
Grounding
Rings or
Lining
Protectors