99
Reference Manual
00809-0300-4750, Rev CA
Advanced Configuration Functionality
March 2016
Advanced Configuration Functionality
Total limit
Configure the parameters that will determine the criteria for activating a HART alert if the
measured net total falls within a set of configured criteria. This functionality can be used for
operating simple batching operations or generating alerts when certain localized values are
met. This parameter can be configured as a discrete output if the transmitter was ordered
with auxiliary outputs enabled (option code AX). If a digital output is configured for
total
limit
, the digital output will activate when the conditions for
total mode
are met.
Total control
This parameter turns the total limit HART alert
ON
or
OFF
.
ON
- The transmitter will generate a HART alert when the defined conditions are met.
OFF
- The transmitter will not generate a HART alert for the total limit.
Total mode
The
total mode
parameter sets the conditions under which the total limit HART alert will
activate. High and low limits exist for each channel and can be configured independently.
> High limit
- The HART alert will activate when the totalizer value exceeds the
high limit
set
point.
< Low limit
- The HART alert will activate when the totalizer value falls below the
low limit
set point.
In range
- The HART alert will activate when the totalizer value is between the
high limit
and
low limit
set points.
Out of range
- The HART alert will activate when the totalizer value exceeds the
high limit
set point or falls below the
low limit
set point.
LOI menu path
Field Mount: Detailed Setup, Output Config, Totalizer, Total Limit
Wall Mount: AUX. FUNCTION
Traditional Fast Keys
1, 4, 2, 3, 5
Device dashboard
2, 2, 4, 5
LOI menu path
Field Mount: Detailed Setup, Output Config, Totalizer, Total Limit,
Total Control
Wall Mount: AUX. FUNCTION
Traditional Fast Keys
1, 4, 2, 3, 5, 1
Device dashboard
2, 2, 4, 5, 4
LOI menu path
Field Mount: Detailed Setup, Output Config, Totalizer, Total Limit, Total
Mode
Wall Mount: AUX. FUNCTION
Traditional Fast Keys
1, 4, 2, 3, 5, 2
Device dashboard
2, 2, 4, 5, 3
Summary of Contents for Rosemount 8750W
Page 2: ......
Page 96: ...84 Reference Manual 00809 0300 4750 Rev CA Operation March 2016 Operation ...
Page 168: ...156 Reference Manual 00809 0300 4750 Rev CA Maintenance March 2016 Maintenance ...
Page 214: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 202 Wiring Diagrams ...
Page 218: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 206 Wiring Diagrams ...
Page 219: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 207 Wiring Diagrams ...
Page 220: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 208 Wiring Diagrams ...
Page 222: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 210 Wiring Diagrams ...
Page 223: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 211 Wiring Diagrams ...
Page 224: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 212 Wiring Diagrams ...
Page 225: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 213 Wiring Diagrams ...
Page 227: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 215 Wiring Diagrams ...
Page 228: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 216 Wiring Diagrams ...
Page 229: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 217 Wiring Diagrams ...
Page 263: ......