Quick Installation and Start-Up
March 2016
Quick Installation and Start-Up
2.3
Transmitter symbols
Caution symbol — check product documentation for details
Protective conductor (grounding) terminal
Failure to follow these installation guidelines could result in death or serious
injury.
Installation and servicing instructions are for use by qualified personnel only. Do
not perform any servicing other than that contained in the operating instructions,
unless qualified.
Verify the installation is done safely and is consistent with the operating
environment.
Ensure the device certification and installation techniques are suitable for the
installation environment.
Explosion hazard. Do not disconnect equipment when a flammable or
combustible atmosphere is present.
To prevent ignition of flammable or combustible atmospheres, disconnect power
before servicing circuits.
Do not connect a Rosemount 8750W Transmitter to a non-Rosemount sensor
that is located in an explosive atmosphere.
Follow national, local, and plant standards to properly earth ground the
transmitter and sensor. The earth ground must be separate from the process
reference ground.
Rosemount Magnetic Flowmeters ordered with non-standard paint options or
non-metallic labels may be subject to electrostatic discharge. To avoid
electrostatic charge build-up, do not rub the flowmeter with a dry cloth or clean
with solvents.
NOTICE
The sensor liner is vulnerable to handling damage. Never place anything through the
sensor for the purpose of lifting or gaining leverage. Liner damage may render the
sensor inoperable.
Metallic or spiral-wound gaskets should not be used as they will damage the liner
face of the sensor.
Correct flange bolt tightening is crucial for proper sensor operation and life. All bolts
must be tightened in the proper sequence to the specified torque specifications.
Failure to observe these instructions could result in severe damage to the sensor
lining and possible sensor replacement.
In cases where high voltage/high current are present near the meter installation,
ensure that proper protection methods are followed to prevent stray voltage /
current from passing through the meter. Failure to adequately protect the meter
could result in damage to the transmitter and lead to meter failure.
Completely remove all electrical connections from both sensor and transmitter
prior to welding on the pipe. For maximum protection of the sensor, consider
removing it from the pipeline.
Summary of Contents for Rosemount 8750W
Page 2: ......
Page 96: ...84 Reference Manual 00809 0300 4750 Rev CA Operation March 2016 Operation ...
Page 168: ...156 Reference Manual 00809 0300 4750 Rev CA Maintenance March 2016 Maintenance ...
Page 214: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 202 Wiring Diagrams ...
Page 218: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 206 Wiring Diagrams ...
Page 219: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 207 Wiring Diagrams ...
Page 220: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 208 Wiring Diagrams ...
Page 222: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 210 Wiring Diagrams ...
Page 223: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 211 Wiring Diagrams ...
Page 224: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 212 Wiring Diagrams ...
Page 225: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 213 Wiring Diagrams ...
Page 227: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 215 Wiring Diagrams ...
Page 228: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 216 Wiring Diagrams ...
Page 229: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 217 Wiring Diagrams ...
Page 263: ......