Rosemount 8700 Series
27
North American Certifications
Factory Mutual (FM)
N0
Non-incendive for Class I, Division 2, Groups A, B, C, and D non-flammable fluids
(8705/8711 T5 at 60 °C; 8707 T3C at 60 °C), and Dust-ignition proof Class II/III,
Division 1, Groups E, F, and G (8705/8711 T6 at 60 °C; 8707 T3C at 60 °C) Hazardous
locations; Enclosure Type 4X
N0
8721 Hygienic Sensor
Factory Mutual (FM) Ordinary Location;
CE Marking; 3-A Symbol Authorization #1222;
EHEDG Type EL
N5
Non-incendive for Class I, Division 2, Groups A, B, C, and D; with intrinsically safe
electrodes for use on flammable fluids (8705/8711 T5 at 60 °C; 8707 T3C at 60 °C),
and Dust-ignition proof Class II/III, Division 1, Groups E, F, and G (8705/8711 T6 at 60
°C; 8707 T3C at 60 °C) Hazardous locations; Enclosure Type 4X
E5
Explosion proof for Class I, Division 1, Groups C and D (8705/8711 T6 at 60 °C), and
Dust-ignition proof Class II/III, Division 1, Groups E, F, and G (8705/8711 T6 at 60 °C),
and non-incendive for Class I, Division 2, Groups A, B, C, and D flammable fluids
(8705/8711 T5 at 60 °C) Hazardous locations; Enclosure Type 4X
Canadian Standards Association (CSA)
N0
Non-incendive for Class I, Division 2, Groups A, B, C, and D non-flammable fluids
(8705/8711 T5 at 60 °C; 8707 T3C at 60 °C), and Dust-ignition proof Class II/III,
Division 1, Groups E, F, and G (8705/8711 T6 at 60 °C; 8707 T3C at 60 °C) Hazardous
locations; Enclosure Type 4X
N0
8721 Hygienic Sensor
Canadian Standards Association (CSA) Ordinary Location;
CE Marking; 3-A Symbol Authorization #1222;
EHEDG Type EL
European Certifications
ND
ATEX Dust Certificate No.: KEMA 06ATEX0006
II 1D Ex tD A20 IP6x T105°C (-50
≤
T
a
≤
65 °C)
0575
Installation Instructions
The cable and conduit entry devices and blanking elements shall be of a certified IP6x
type, suitable for the conditions of use and correctly installed. At maximum ambient
temperatures or at process temperatures above 60 °C heat resistant cables with a
temperature rating of at least 90 °C shall be used.
The surface temperature of 105 °C is based on a maximum ambient temperature of
65 °C. When the process temperature is higher than the maximum ambient
temperature (up to a maximum of 180 °C), the surface temperature will be the process
temperature plus 40 °K.