37
Reference Manual
00809-0200-4705, Rev AB
May 2017
5.4
Replacing the power module
When the power module needs to be replaced, remove the power module cover and the depleted power
module. Replace the power module with a new Black Power Module, SmartPower
™
Solutions model
number 701PBKKF (part number: 00753-9220-0001). Then replace the cover and tighten it to
specification. Always ensure a proper seal so metal touches metal, but do not over-tighten. After the
power module has been replaced, navigate to AMS Device Manager to reset the power module function.
This also can be done via the handheld
3, 5, 1, 3
.
Handling the power module
The black power module with the wireless unit contains two “C” size primary lithium/thionyl chloride
batteries. Each battery contains approximately 2.5 grams of lithium, for a total of 5 grams in each pack.
Under normal conditions, the battery materials are self-contained and are not reactive as long as the
batteries and the battery pack integrity are maintained. Care should be taken to prevent thermal,
electrical or mechanical damage. Contacts should be protected to prevent premature discharge.
Use caution when handling the power module, it may be damaged if dropped from heights in excess of
20 ft.
Battery hazards remain when cells are discharged.
Environmental considerations
As with any battery, local environmental rules and regulations should be consulted for proper
management of spent batteries. If no specific requirements exist, recycling through a qualified recycler is
encouraged. Consult the materials safety data sheet for battery specific information.
Shipping considerations
The unit was shipped to you without the power module installed. Remove the power module prior to
shipping.
Each power module contains two “C” size primary lithium batteries. Primary lithium batteries are
regulated in transportation by the U.S. Department of Transportation, and are also covered by
International Air Transport Association (IATA), International Civil Aviation Organization (ICAO), and
European Ground Transportation of Dangerous Goods (ARD). It is the responsibility of the shipper to
ensure compliance with these or any other local requirements. Consult current regulations and
requirements before shipping.
When replacing the power module, the device will stop totalizing data. The device will resume from
where the last total was periodically recorded (within a five minute window). Example: The best practice
when replacing the power module is to make note of the totalized volume before removing the power
module. After installing a new power module, the totalized volume can be set to the previous recorded
value.
Summary of Contents for Rosemount 705
Page 2: ......
Page 6: ...vi Reference Manual 00809 0200 4705 Rev AB Contents May 2017 Contents ...
Page 10: ...2 Reference Manual 00809 0200 4705 Rev AB Introduction May 2017 Introduction ...
Page 32: ...28 Reference Manual 00809 0200 4705 Rev AB Commissioning May 2017 Commissioning ...
Page 69: ......