99
Reference Manual
00809-0200-4728, Rev RA
Appendix A: Specifications and Reference Data
April 2015
Specifications and Reference Data
Cu 50 (
α
= 0.00428)
GOST 6651-94
–185 to 200
–301 to 392
10
18
±1.34
± 2.41
±0.03% of span
Cu 100 (
α
= 0.00426)
GOST 6651-94
–50 to 200
–58 to 392
10
18
±0.67
± 1.20
±0.03% of span
Cu 100 (
α
= 0.00428)
GOST 6651-94
–185 to 200
–301 to 392
10
18
±0.67
± 1.20
±0.03% of span
Thermocouples
(4)
Type B
(5)
NIST Monograph 175, IEC 584
100 to 1820
212 to 3308
25
45
± 0.77
± 1.39
±0.03% of span
Type E
NIST Monograph 175, IEC 584
–50 to 1000
–58 to 1832
25
45
± 0.20
± 0.36
±0.03% of span
Type J
NIST Monograph 175, IEC 584
–180 to 760
–292 to 1400
25
45
± 0.35
± 0.63
±0.03% of span
Type K
(6)
NIST Monograph 175, IEC 584
–180 to 1372
–292 to 2501
25
45
± 0.50
± 0.90
±0.03% of span
Type N
NIST Monograph 175, IEC 584
–200 to 1300
–328 to 2372
25
45
± 0.50
± 0.90
±0.03% of span
Type R
NIST Monograph 175, IEC 584
0 to 1768
32 to 3214
25
45
± 0.75
± 1.35
±0.03% of span
Type S
NIST Monograph 175, IEC 584
0 to 1768
32 to 3214
25
45
± 0.70
± 1.26
±0.03% of span
Type T
NIST Monograph 175, IEC 584
–200 to 400
–328 to 752
25
45
± 0.35
± 0.63
±0.03% of span
DIN Type L
DIN 43710
–200 to 900
–328 to 1652
25
45
± 0.35
± 0.63
±0.03% of span
DIN Type U
DIN 43710
–200 to 900
–328 to 1112
25
45
± 0.35
± 0.63
±0.03% of span
Type W5Re/W26Re
ASTM E 988-96
0 to 2000
32 to 3632
25
45
± 0.70
± 1.26
±0.03% of span
GOST Type L
GOST R 8.585-2001
–200 to 800
–328 to 1472
25
45
± 1.00
± 1.26
±0.03% of span
Other input types
Millivolt Input
–10 to 100 mV
±0.015 mV
±0.03% of span
2-, 3-, 4-wire Ohm Input
0 to 2000 ohms
±0.45 ohm
±0.03% of span
(1) No minimum or maximum span restrictions within the input ranges. Recommended minimum span will hold noise within accuracy specification with
damping at zero seconds.
(2) The published digital accuracy applies over the entire sensor input range. Digital output can be accessed by HART or F
OUNDATION
fieldbus Communications
or Rosemount control system.
(3) Total Analog accuracy is the sum of digital and D/A accuracies. This is not applicable for F
OUNDATION
fieldbus.
(4) Total digital accuracy for thermocouple measurement: sum of digital ac0.5 °C. (cold junction accuracy).
(5) Digital accuracy for NIST Type B T/C is ±3.0 °C (±5.4 °F) from 100 to 300 °C (212 to 572 °F).
(6) Digital accuracy for NIST Type K T/C is ±0.70 °C (±1.26 °F) from –180 to –90 °C (–292 to –130 °F).
Table A-2. Rosemount 644 Input Options and Accuracy
Sensor options
Sensor reference
Input ranges
Recommended
min. span
(1)
Digital accuracy
(2)
D/A accuracy
(3)
Summary of Contents for Rosemount 644
Page 2: ......
Page 4: ...iv Reference Manual 00809 0200 4728 Rev RA Title Page April 2015 ...
Page 10: ...x Reference Manual 00809 0200 4728 Rev RA Table of Contents April 2015 Table of Contents ...
Page 180: ...Index 2 Reference Manual 00809 0200 4728 Rev RA Index April 2015 Index ...
Page 181: ......