Quick Installation Guide
00825-0200-4585, Rev AA
January 2009
Flanged Flo-Tap 585 Annubar
11
S
TEP
7: I
NSERT
THE
A
NNUBAR
1. Open the isolation valve fully
.
2. Rotate the crank clockwise. If a power drill with an adapter is used, do not exceed 200
revolutions per minute.
3. Continue rotating the crank until the sensor firmly contacts the opposite side of the pipe.
a. The orange stripes are visual indication of when the sensor is approaching the
opposite side wall.
b. As the orange stripes approach the support plate, remove the power drill and continue
cranking manually. Place a finger above the packing gland while cranking. Vibration
and movement will occur. When vibration and movements stop, the sensor is in
contact with the opposite side wall.
NOTE
Do not place finger above packing gland for high temperature applications.
c. Turn the handle an additional
1
/
4
to
1
/
2
turn to secure the sensor.
Figure 10. Insert the Sensor
Drive Lock Pin