![Emerson Rosemount 5408 Reference Manual Download Page 123](http://html.mh-extra.com/html/emerson/rosemount-5408/rosemount-5408_reference-manual_103670123.webp)
Suppress false echoes using AMS Device Manager
Procedure
1. Select
Service Tools
→
Echo Tuning
→
Suppress
. Suppressed echoes are shown in
the table.
2. Select
Suppress
or
Remove Suppression
.
3. Type the distance to the echo that should be added to or removed from the list, and
then select
Next
.
4. Select
Finish
when the
Method Complete
message appears.
Suppress false echoes using handheld communicator
Procedure
1. Select
Service Tools
→
Echo Tuning
→
Suppress
.
2. To view the current suppressed echoes, select
Suppressed Echoes
.
3. Select
Suppress
or
Remove Suppression
.
4. Type the distance to the echo that should be added to or removed from the list, and
then select
Enter
.
7.4.3
Perform an analog loop test
Perform an analog loop test using Rosemount Radar Master
Plus
During a loop test, the transmitter outputs a fixed value (4 mA, 20 mA, or user-selected
value). The loop test command verifies the output of the transmitter, the integrity of the
loop, and the operations of any recorders or similar devices installed in the loop. A loop
test can also be used to determine the need for an analog output calibration.
Prerequisites
If the transmitter is equipped with a TEST terminal, current can be measured directly at the
terminal block without disconnecting any signal wires.
Procedure
1. Under
Service Tools
, select
Simulate
.
2. Select
Loop Test
.
3. Select
4 mA
,
20 mA
, or
Other
.
4. If
Other
is selected, enter the desired value.
5. Select
Start
.
6. Measure the loop current.
7. Select
Stop
to end loop test.
Related information
Calibrate analog out using Rosemount Radar Master Plus
Reference Manual
Service and troubleshooting
00809-0100-4408
February 2020
Rosemount 5408 and 5408:SIS Level Transmitters
123
Summary of Contents for Rosemount 5408
Page 4: ...4 ...
Page 8: ...Contents Reference Manual February 2020 00809 0100 4408 viii Reference Manual ...
Page 10: ...Introduction Reference Manual February 2020 00809 0100 4408 10 Reference Manual ...
Page 20: ...Transmitter overview Reference Manual February 2020 00809 0100 4408 20 Reference Manual ...
Page 72: ...Mechanical installation Reference Manual February 2020 00809 0100 4408 72 Reference Manual ...
Page 100: ...Operation Reference Manual February 2020 00809 0100 4408 100 Reference Manual ...
Page 240: ...Product certifications Reference Manual February 2020 00809 0100 4408 240 Reference Manual ...
Page 242: ...Product certifications Reference Manual February 2020 00809 0100 4408 242 Reference Manual ...
Page 244: ...Product certifications Reference Manual February 2020 00809 0100 4408 244 Reference Manual ...
Page 246: ...Product certifications Reference Manual February 2020 00809 0100 4408 246 Reference Manual ...
Page 248: ...Product certifications Reference Manual February 2020 00809 0100 4408 248 Reference Manual ...