![Emerson Rosemount 5408 Reference Manual Download Page 113](http://html.mh-extra.com/html/emerson/rosemount-5408/rosemount-5408_reference-manual_103670113.webp)
Table 7-7: Troubleshooting the 4-20 mA/HART Output
(continued)
Symptom
Recommended actions
Transmitter will not respond to changes in level.
•
Verify level is between the 4 and 20 mA set points.
•
Verify output is not in alarm condition.
•
Verify transmitter is not in loop test or simulation mode.
There is no HART communication (lost device
communication).
•
Verify power supply voltage is adequate at signal terminals,
see
•
Check load resistance (250 ohms minimum).
•
Check if transmitter is at an alternate HART address.
•
Check current analog output value to verify that transmitter
hardware works.
•
Verify the blue plug is attached to the TEST terminal (if
applicable). When unplugged, HART communication to
configuration tool may be compromised.
7.4
Service and troubleshooting tools
This section briefly describes tools and functions in the Rosemount Radar Master Plus,
AMS Device Manager, and handheld communicator, which may be useful for service and
troubleshooting of the Rosemount 5408 and 5408:SIS Level Transmitters.
7.4.1
Using the echo curve
The Rosemount Radar Master Plus software includes functions for viewing and recording
single instances or movies of the echo curve. The echo curve represents the tank, as seen
by the radar transmitter. Each peak corresponds to a strong reflection of the radar signal.
When connected to Rosemount Radar Master Plus, past measurement records and echo
curves including the 10 highest peaks, as well as the 50 last alert events are automatically
transferred from the transmitter’s internal memory to the hard drive on your local
computer. Past measurement records are then available the next time you connect to the
transmitter using the level trend timeline.
Reference Manual
Service and troubleshooting
00809-0100-4408
February 2020
Rosemount 5408 and 5408:SIS Level Transmitters
113
Summary of Contents for Rosemount 5408
Page 4: ...4 ...
Page 8: ...Contents Reference Manual February 2020 00809 0100 4408 viii Reference Manual ...
Page 10: ...Introduction Reference Manual February 2020 00809 0100 4408 10 Reference Manual ...
Page 20: ...Transmitter overview Reference Manual February 2020 00809 0100 4408 20 Reference Manual ...
Page 72: ...Mechanical installation Reference Manual February 2020 00809 0100 4408 72 Reference Manual ...
Page 100: ...Operation Reference Manual February 2020 00809 0100 4408 100 Reference Manual ...
Page 240: ...Product certifications Reference Manual February 2020 00809 0100 4408 240 Reference Manual ...
Page 242: ...Product certifications Reference Manual February 2020 00809 0100 4408 242 Reference Manual ...
Page 244: ...Product certifications Reference Manual February 2020 00809 0100 4408 244 Reference Manual ...
Page 246: ...Product certifications Reference Manual February 2020 00809 0100 4408 246 Reference Manual ...
Page 248: ...Product certifications Reference Manual February 2020 00809 0100 4408 248 Reference Manual ...