SECTION 8.0
CALIBRATION — TEMPERATURE
2. Press CAL on the remote controller.
3. Press NEXt until the
tEMP AdJ
submenu appears. Press Enter.
4. the
tEMP
prompt appears. use the arrow keys to change the display to
match the temperature measured using the standard thermometer.
Press ENtER to save.
5. the
tEMP AdJ
sub-menu appears. Press RESEt to return to the main
display.
8.2. PROCEDURE USING THE INFRARED REMOTE CONTROLLER
1. Place the sensor and a calibrated reference thermometer in a container of water at ambient temperature. Be
sure the temperature element in the sensor is completely submerged by keeping the sensor tip at least three
inches below the water level. Stir continuously. Allow at least 20 minutes for the standard thermometer, sen-
sor, and water to reach constant temperature.
8.3. PROCEDURE USING Deltav
1. Place the sensor and a calibrated reference thermometer in a container of water at ambient temperature. Be
sure the temperature element in the sensor is completely submerged by keeping the sensor tip at least three
inches below the water level. Stir continuously. Allow at least 20 minutes for the standard thermometer, sen-
sor, and water to reach constant temperature.
2. Access: deltaV Explorer/Context Menu
Standardize temperature (method_sv_cal)
Method Steps:
a. Is temperature stable?: Yes; No; Abort
b. If yes is chosen, enter the new temperature value.