51
MODEL 5081-A
SECTION 7.0
PROGRAMMING
7.6.2 Procedure using the infrared remote controller.
1. Press PRoG on the remote controller.
2. Press NEXt until the
LinE FrEq
submenu appears. Press ENtER.
3. use
é
or
ê
to toggle the line frequency between
50
and
60
Hz. Press ENtER to
save.
4. Press RESEt to return to the main display.
7.6 LINE FREQUENCy
7.6.1 Purpose
this section describes how to maximize noise rejection by entering the frequency of the mains power into the transmitter.
PRoGRAM
line fre9
EXIt
NEXt ENtER
PRoGRAM
line
EXIt
ENtER
60