6. Set the stabilization range to between 0.01 and 9.99 ppm. the default values are
shown in the table. Press ENtER to save.
7. the display returns to the
StABiLiSE
prompt. Press NEXt. the next screen depends
on the measurement being made. For free or total chlorine see step 8. For oxygen,
see step 9. For ozone see step 10.
8. If the measurement is free or total chlorine, the
SLOPE
prompt appears. use
é
or
ê
to toggle between
SnGL
(single) or
duAL
(dual) slope. Press ENtER. Go to step 10.
NOTE
For the vast majority of applications, single slope calibration is accept-
able. Dual slope calibration is useful in fewer than 5 % of applications.
9. If the measurement is oxygen, the
SALnty
(salinity) prompt appears. use the arrow
keys to enter the salinity of the water. Press ENtER. Go to step 10.
10. the display returns to the
SPAn CAL
screen. Press NEXt.
11. the
0 CAL
screen appears. Press ENtER.
12. Enter the desired zero limit. the units are the same as the units programmed in Sec-
tion 7.5. default limits are given in the table.
13. Press RESEt to return to the main display.
PRoGRAM
delta
EXIt
ENtER
0
. 0 5
PRoGRAM
Stabilise
EXIt
NEXt ENtER
PRoGRAM
salnty
EXIt
ENtER
0
. 0 0
PRoGRAM
slope
EXIt
ENtER
SngL
PRoGRAM
span Cal
EXIt
NEXt ENtER
PRoGRAM
0 Cal
EXIt
NEXt ENtER
PRoGRAM
limit
EXIt
ENtER
0
0
.00
oxygen
0.05 ppm or 1%
Free chlorine
0.05 ppm
total chlorine
0.05 ppm
ozone
0.01 ppm
oxygen (ppm)
0.05 ppm
oxygen (ppb)
2.0 ppb
oxygen (% saturation)
1%
Free chlorine
0.05 ppm
total chlorine
0.05 ppm
ozone
0.01 ppm or 10 ppb
MODEL 5081-A
SECTION 7.0
PROGRAMMING
49