33
MODEL 5081-A
SECTION 5.0
DISPLAy AND OPERATION WITH INFRARED REMOTE CONTROLLER
5.3 MENU TREE
the Model 5081-A transmitter has three menus: CALIBRAtE, PRoGRAM, and dIAGNoSE. under the Calibrate and
Program menus are several submenus. under each submenu are a number of prompts. the dIAGNoSE menu shows
the reader diagnostic variables that are useful in troubleshooting. Figure 5-4, on the following page, shows the complete
menu tree.
5.4 DIAGNOSTIC MESSAGES
Whenever a warning or fault limit has been exceeded, the transmitter displays diagnostic fault messages. the display
alternates between the main display and the diagnostic message. See Section 15.0 for the meaning of fault and warning
messages.
5.5 SECURITy
5.5.1 Purpose.
use the security code to prevent program settings and calibrations from accidentally being changed.
to program a security code, refer to Section 7.5.
1. If settings are protected with a security code, pressing PRoG or CAL on the remote
controller causes the
Id
screen to appear.
2. use the arrow keys to enter the security code. Press ENtER.
3. If the security code is correct, the first submenu appears. If the code is incorrect, the
process display reappears.
4. to retrieve a forgotten code number, enter 555 at the
Id
prompt. the present security
code will appear.
PRoGRAM
Id
EXIt
ENtER
0
00
5.5.2 Change security code using Fieldbus.
Access: deltaV Explorer/transducer Block/Properties Identification tab
Parameter: Security Code for Infrared Remote (LoCAL_oPERAtoR_INtERFACE_tAG)
Enter desired security code (0 - 999)
5.6 USING HOLD
during calibration, the sensor may be exposed to solutions having concentration outside the normal range of the process.
to prevent false alarms and undesired operation of chemical dosing pumps, place the transmitter in hold during calibra-
tion. Activating hold keeps the transmitter output at the last value or sends the output to a previously determined value.
See Section 7.3, output Ranging, for details.
After calibration, reinstall the sensor in the process stream. Wait until readings have stabilized before deactivating Hold.
to activate or deactivate Hold:
1. Press HoLd on the remote controller.
2. the
HoLd
prompt appears in the display. Press
é
or
ê
to toggle Hold between
On
and
OFF
.
3. Press ENtER to save.