Figure 2-2: In-line Protected Gage Vent Low Side Pressure Port
Aluminum
Polished 316 SST
A. Low side pressure port (atmospheric reference)
Clamping
When installing clamp, follow recommended torque values provided by
gasket manufacturer.
Note
To maintain performance, torquing a 1.5-in. Tri Clamp beyond 50 in-lb. is
not recommended on pressure ranges below 20 psi.
June 2020
Quick Start Guide
Quick Start Guide