55
FD_FAIL_ALM
This parameter is used primarily to broadcast a change in the associated active
conditions, which are not masked, for this alarm category to a Host System.
56
FD_OFFSPEC_ALM
57
FD_MAINT_ALM
58
FD_CHECK_ALM
59
FD_FAIL_PRI
This parameter allows the user to specify the priority of this alarm category.
60
FD_OFFSPEC_PRI
61
FD_MAINT_PRI
62
FD_CHECK_PRI
63
FD_SIMULATE
This parameter allows the conditions to be manually supplied when simulation is
enabled. When simulation is disabled both the diagnostic simulate value and the
diagnostic value tracks the actual conditions. The simulate jumper is required for
simulation to be enabled and while simulation is enabled the recommended action
will show that simulation is active.
64
FD_RECOMMEN_ACT
This parameter is a device enumerated summarization of the most severe
condition or conditions detected. The DD help should describe by enumerated
action, what should be done to alleviate the condition or conditions. 0 is defined as
Not Initialized, 1 is defined as No Action Required, all others defined by
manufacturer.
65
FD_EXTENDED_ACTIVE
An optional parameter or parameters to allow the user finer detail on conditions
causing an active condition in the FD_*_ACTIVE parameters.
66
FD_EXTENDED_MAP
An optional parameter or parameters to allow the user finer control on enabling
conditions contributing to the conditions in FD_*_ACTIVE parameters.
67
COMPATIBILITY_REV
This parameter is used when replacing field devices. The correct value of this
parameter is the DEV_REV value of the replaced device.
68
HARDWARE_REVISION
Hardware revision.
69
SOFTWARE_REV
Software revision of source code with resource block.
70
PD_TAG
PD tag description of device.
71
DEV_STRING
This is used to load new licensing into the device. The value can be written but will
always read back with a value of 0.
72
DEV_OPTIONS
Indicates which miscellaneous device licensing options are enabled.
73
OUTPUT_BOARD_SN
Output board serial number. For the Rosemount 2230 Display this is the same as
Main Label Device ID which can be found on the main label that is attached to the
housing.
74
FINAL_ASSY_NUM
Final assembly number given by manufacturer.
75
DOWNLOAD_MODE
Gives access to the boot block code for over-the-wire downloads.
0 = Uninitialized
1 = Run mode
2 = Download mode
76
HEALTH_INDEX
Parameter representing the overall health of the device, 100 being perfect and 1
being non-functioning. The value is based on the active PWA alarms.
77
FAILED_PRI
Designates the alarming priority of the FAILED_ALM and also used as switch b/w FD
and legacy PWA. If value is greater than or equal to 1 then PWA alerts will be active
in device else device will have FD alerts.
78
RECOMMENDED_ACTION
Enumerated list of recommended actions displayed with a device alert.
Index
Number
Parameter
Description
Summary of Contents for Rosemount 2230
Page 1: ...Reference Manual 00809 0100 2230 Rev CA July 2017 Rosemount 2230 Graphical Field Display ...
Page 2: ......
Page 4: ...ii Reference Manual 00809 0100 2230 Rev CA Title Page July 2017 Title Page ...
Page 16: ...6 Reference Manual 00809 0100 2230 Rev CA Introduction July 2017 Introduction ...
Page 28: ...18 Reference Manual 00809 0100 2230 Rev CA Overview July 2017 Overview ...
Page 44: ...34 Reference Manual 00809 0100 2230 Rev CA Installation July 2017 Installation ...
Page 154: ...144 Reference Manual 00809 0100 2230 Rev CA Index July 2017 Index ...
Page 155: ......