cloth. If paint is ordered through a special option code, contact the
manufacturer for more information.
3. The device contains a thin wall diaphragm less than 1 mm thickness
that forms a boundary between zone 0 (process connection) and zone
1 (all other parts of the equipment). The model code and datasheet
are to be consulted for details of the diaphragm material. Installation,
maintenance and use shall take into account the environmental
conditions to which the diaphragm shall be subjected. The
manufacturer's instructions for installation and maintenance shall be
followed in detail to assure safety during its expected lifetime.
4. Flameproof joints are not intended for repair.
I1 ATEX Intrinsic Safety
Certificate
Baseefa08ATEX0129X
Standards
EN60079-0:2012+A11:2013, EN60079-11:2012
Markings
II 1 G Ex ia IIC T4 Ga (–60 °C ≤ T
a
≤ +70 °C)
Table 7-2: Input Parameters
Input parameter
HART
®
Fieldbus/PROFIBUS
®
Voltage U
i
30 V
30 V
Current I
i
200 mA
300 mA
Power P
i
1 W
1.3 W
Capacitance C
i
0.012 μF
0 μF
Inductance L
i
0 mH
0 mH
Special Conditions for Safe Use (X):
1. If the equipment is fitted with an optional 90V transient suppressor, it
is incapable of withstanding the 500V isolation from earth test and this
must be taken into account during installation.
2. The enclosure may be made of aluminum alloy and given a protective
polyurethane paint finish; however care should be taken to protect it
from impact and abrasion when located in Zone 0.
IA ATEX FISCO
Certificate
Baseefa08ATEX0129X
Standards
EN60079-0:2012+A11:2013, EN60079-11:2012
Markings
II 1 G Ex ia IIC T4 Ga (–60 °C ≤ T
a
≤ +60 °C)
Quick Start Guide