52
Reference Manual
00809-0100-4100, Rev AA
Section 5: Operation and Maintenance
June 2012
Table 5-1. Recommended Calibration Tasks
Note:
A Field Communicator or AMS is required for all sensor and output trim procedures.
Rosemount 3051C Range 4 and Range 5 transmitters require a special calibration procedure
when used in differential pressure applications under high static line pressure.
5.2.1
Sensor Trim Overview
Trim the sensor using either sensor or zero trim functions. Trim functions vary in complexity and
are application-dependent. Both trim functions alter the transmitter’s interpretation of the
input signal.
Zero trim is a single-point offset adjustment. It is useful for compensating for mounting position
effects and is most effective when performed with the transmitter installed in its final mounting
position. Since this correction maintains the slope of the characterization curve, it should not be
used in place of a sensor trim over the full sensor range.
When performing a zero trim, ensure that the equalizing valve is open and all wet legs are filled
to the correct levels.
Note
Do not perform a zero trim on 3051 Wireless absolute pressure transmitters. Zero trim is zero
based, and absolute pressure transmitters reference absolute zero. To correct mounting
position effects on a 3051 Wireless absolute pressure transmitter, perform a low trim within the
sensor trim function. The low trim function provides an offset correction similar to the zero trim
function, but it does not require zero-based input.
Transmitter
Bench Calibration Tasks
Field Calibration Tasks
3051CD
3051CG
3051L
3051TG, Range
1-4
1.
Set output configuration parameters:
a.
Set the range points.
b.
Set the output units.
c.
Set the output type.
d.
Set the damping value.
2.
Optional
: Perform a sensor trim.
(Accurate pressure source required.)
1.
Reconfigure parameters if
necessary.
2.
Zero trim the transmitter to
compensate for mounting effects
or static pressure effects.
3.
Optional
: Perform an analog output
trim. (Accurate multimeter
required)
3051CA
3051TA
3051TG, Range 5
1.
Set output configuration parameters:
a.
Set the range points.
b.
Set the output units.
c.
Set the output type.
d.
Set the damping value.
2.
Optional
: Perform a sensor trim if
equipment available (accurate
absolute pressure source required),
otherwise perform the low trim value
section of the sensor trim procedure.
1.
Reconfigure parameters if
necessary.
2.
Perform low trim value section of
the sensor trim procedure to
correct for mounting position
effects.
3.
Optional
: Perform an analog output
trim (Accurate multimeter
required)
Summary of Contents for Rosemount 2051 Wireless Series
Page 2: ......
Page 12: ...4 Reference Manual 00809 0100 4102 Rev AA Section 1 Introduction June 2012 Introduction ...
Page 32: ...24 Reference Manual 00809 0100 4102 Rev AA Section 2 Configuration June 2012 Configuration ...
Page 110: ......
Page 113: ......
Page 117: ...iv Reference Manual 00809 0100 4100 Rev AA Title Page June 2012 ...
Page 123: ...x Reference Manual 00809 0100 4100 Rev AA Table of Contents June 2012 Table of Contents ...
Page 167: ...44 Reference Manual 00809 0100 4100 Rev AA Section 3 Installation June 2012 Installation ...
Page 168: ...45 Reference Manual 00809 0100 4100 Rev AA Section 3 Installation June 2012 Installation ...
Page 169: ...46 Reference Manual 00809 0100 4100 Rev AA Section 3 Installation June 2012 Installation ...
Page 249: ......