Section 7:
Calibrate
7.1
Calibration procedure
The calibration procedure requires about five minutes to complete. If
gas is not applied at the appropriate time, a calibration failure may
occur. Refer to
For accurate performance, the Millennium should be calibrated using
50% span of the specific gas of concern. The concentration of gas,
corresponding to 100% of full scale, is converted to a linear
4–20 mA output signal which can be powered from the primary dc
supply of the instrument. Power up the unit for at least 4 hours BEFORE
the first calibration.
The following calibration procedure should be followed to ensure an
accurate correlation between the 4–20 mA output signal and the gas
concentration.
To compensate for distance when remotely calibrating (sensor wired
for separation), decrease the tubing diameter or increase the
calibration gas flow rate between canister and sensor. Always confirm
calibration by applying gas directly at the sensor.
Figure 7-1 Calibration Procedure
Summary of Contents for Net Safety MLP-A-SC1100
Page 2: ......
Page 10: ......
Page 12: ...Operations Manual January 2017 FGD MAN 0047 2 ...
Page 22: ...Operations Manual January 2017 FGD MAN 0047 12 ...
Page 26: ...Operations Manual January 2017 FGD MAN 0047 16 ...
Page 28: ...Operations Manual January 2017 FGD MAN 0047 18 ...
Page 32: ...Operations Manual January 2017 FGD MAN 0047 22 ...
Page 38: ...Operations Manual January 2017 FGD MAN 0047 28 ...
Page 45: ...Operations Manual FGD MAN 0047 January 2017 35 ...
Page 47: ...Operations Manual FGD MAN 0047 January 2017 37 ...
Page 51: ...Operations Manual FGD MAN 0047 January 2017 41 ...
Page 53: ...Notes ...
Page 54: ...Notes ...
Page 55: ......