LF-Series Flowmeters: Sensor/Transmitter Installation
59
I/O Wiring – DIN CIO Transmitters
I/
O
Wiring – DIN
CIO
Retu
rn
P
o
li
c
y
Sp
ec
ifi
c
ati
o
n
s
I/
O
Wir
ing – DIN AN
Figure 10-8
Frequency output wiring – Terminals 31 & 32 (Channel C) – External power
10.5
Discrete output wiring
Discrete output wiring depends on whether you are wiring terminals 23 and 24 (Channel B) or
terminals 31 and 32 (Channel C), and also on whether you have configured the terminals for internal
or external power. The following diagrams are examples of proper wiring for these configurations:
•
Channel B (DO1), internal power – Figure 10-9
•
Channel B (DO1), external power – Figure 10-10
•
Channel C (DO2), internal power – Figure 10-11
•
Channel C (DO2), external power – Figure 10-12
3–30 VDC
Pull-up resistor
Counter
Note: Refer to Figure 10-15 for recommended
resistor versus supply voltage.
000042
–
+
+
–
CAUTION
Excessive current will damage the transmitter.
Do not exceed 30 VDC input. Terminal current must be less than 500 mA.
Summary of Contents for Micro Motion LF Series
Page 6: ...iv LF Series Flowmeters Sensor Transmitter Installation ...
Page 10: ...4 LF Series Flowmeters Sensor Transmitter Installation ...
Page 14: ...8 LF Series Flowmeters Sensor Transmitter Installation ...
Page 28: ...22 LF Series Flowmeters Sensor Transmitter Installation ...
Page 36: ...30 LF Series Flowmeters Sensor Transmitter Installation ...
Page 48: ...42 LF Series Flowmeters Sensor Transmitter Installation ...
Page 52: ...46 LF Series Flowmeters Sensor Transmitter Installation ...
Page 58: ...52 LF Series Flowmeters Sensor Transmitter Installation ...
Page 72: ...66 LF Series Flowmeters Sensor Transmitter Installation ...
Page 86: ...80 LF Series Flowmeters Sensor Transmitter Installation ...
Page 102: ...96 LF Series Flowmeters Sensor Transmitter Installation ...
Page 103: ......