Table 9-1: Status LED and device status (MMI mode)
Status LED condition
Device status
Solid green
No alerts are active.
Solid yellow
One or more alerts are active with
Alert Severity
= Out of
Specification, Maintenance Required, or Function Check.
Solid red
One or more alerts are active with
Alert Severity
= Failure.
Flashing yellow (1 Hz)
The
Function Check in Progress
alert is active.
Table 9-2: Network status LED and Ethernet network connection status
Network status LED condition Network status
Flashing green
No connections made with primary protocol host.
Solid green
Connection made with primary protocol host.
Flashing red
Connection from primary protocol host has timed out.
Solid red
Address Conflict Detection (ACD) algorithm has detected a
duplicate IP address.
LCD panel
In normal operation, the LCD panel shows the current value of the display variables, and
their measurement units.
The LCD panel also provides access to the display menus and alert information. From the
display menus, you can:
•
View the current configuration and make configuration changes.
•
Perform procedures such as loop testing and zero verification.
•
Run batches.
The alert information allows you to see which alerts are active, acknowledge the alerts
individually or as a group, and to see more detailed information for individual alerts.
9.1
Access and use the display menus
The display menus allow you to perform most configuration, administration, and
maintenance tasks.
The four switches,
⇦⇧⇩⇨
, are used to navigate the menus, make selections, and enter
data.
Procedure
1. Observe the action bar at the bottom of the LCD panel.
The action bar displays
Menu
⇨
.
2. Press your thumb or finger over the
⇨
membrane switch to activate it.
The top-level menu is displayed.
Components of the transmitter display
Installation Manual
March 2022
00825-0100-1600
40
Micro Motion 1600 Ethernet Transmitters