3.3
Rotate the transmitter on the sensor (optional)
In integral installations, you can rotate the transmitter on the sensor up to 360º in 45º
increments.
Procedure
1. Using a 4 mm hex key, loosen and remove the clamp securing the transmitter head
in place.
Figure 3-3: Removal of the sensor clamp
2. Gently lift the transmitter straight up, and rotate the transmitter to the desired
position.
You can rotate the transmitter to any of the eight positions, but a stop exists that
will not allow a full 360° rotation.
3. Gently lower the transmitter onto the base, confirming that the transmitter is in a
locked position.
4. Replace the clamp in its original position and tighten the cap screw. Torque to 29 in
lbf (3.28 N m) to 31 in lbf (3.50 N m).
3.4
Rotate the transmitter display
Configure the software to rotate the transmitter display 0°, 90°, 180°, or 270°. You cannot
physically rotate the display.
Procedure
1. Choose
Menu
→
Configuration
→
Display Settings
→
Rotation
.
2. Select the appropriate direction.
Installation Manual
Mounting and sensor wiring
00825-0100-1600
March 2022
Installation Manual
15