Instruction Manual
D102748X012
DLC3010 Digital Level Controller
Maintenance & Troubleshooting
May 2018
89
Figure 6‐3. Installing Inner Guide and Access Handle Assembly
ZERO LOCKING PIN
INNER GUIDE
(KEY 11)
VENT HOLE
LUBRICATE
THIS SURFACE
VENT HOLES
HANDLE
ASSEMBLY
(KEY 12)
SCREWS (KEY 13)
LUBRICATE
THIS SURFACE
TRANSDUCER
HOUSING
ACCESS HOLE
E0381
it will contact the transducer housing.
8. Install the handle assembly (key 12) in the slot of the transducer housing over the inner guide (key 11) so that the
vent holes in the handle assembly are over the access hole.
9. Install two screws (key 13) to secure the handle assembly (key 12) to the inner guide (key 11). Tighten the screws
to 0.48 N
S
m (4.2 lbf
S
in).
zero locking pin engages the lever assembly. Also check for free travel of the lever assembly when the handle is in
the unlocked position.
11. Install the coupling shield (key 16) and secure with the two screws (key 13). Tighten the screws to 0.48 N
S
m
(4.2 lbf
S
in).
13. When re‐installing the digital level controller, follow the appropriate procedure outlined in the Installation section.
Also setup the digital level controller as described in the Setup and Calibration section.
Lever Assembly
Removing the Lever Assembly
1. Remove the digital level controller from the sensor as described in Removing the Digital Level Controller from the
Sensor.
2. Loosen and remove the hex nuts (key 34) from the studs (key 33) and remove the adapter ring (key 32).
3. Remove the coupling shield (key 16) by removing the two screws (key 13). Take care not to drop the screws into the
lever assembly compartment where they will be attracted by the magnets.