Instruction Manual
Installation
LIQ_MAN_6081-P
February 2014
Installation
9
2.6 Power Module Installation
The section describes the procedure for installation of the power module (
PN 701PBKKF or
PN 00753-9220-0001
). The power module should stored in a safe place with a controlled
environment until the 6081 is ready for live operation. For first time installation of the power
module, follow these steps:
1. Unscrew the two long machine screws to remove the rear cover of the 6081. Separate the rear
cover from the central housing by manually prying the sections apart. Do not use screwdrivers
or tools to separate these housing parts. The parts are sealed with an o-ring.
Figure 2-3. Removing rear cover
2. Before installation, note the safety warning, disposal instructions and part information on the
connection-side label of the power module.
Figure 2-4. Power Module Warning Label
3. With the 6081 front display section facing away from you, align the power module pack
with the curved surface of the pack facing towards you and the small protruding connector
facing away from you. Make sure to align the power module and its keyed connector with the
connection receptacle in the middle of the instrument’s terminal block area.
Figure 2-5. Installing the Power Module