6.6 Cleaning and Disinfection of the SonoTT
TM
Clamp-On Transducer
Make sure no detergents, disinfectants or other liquids can leak into the plug of the
SonoTT
TM
Clamp-On Transducer.
For the cleaning of the device, we recommend the use of common household detergents. In order to avoid
stains and dirt from drying out, grime residue should be removed after each application.
To clean the SonoTT
TM
Clamp-On Transducer, use a lint-free cloth. Be sure to remove any dirt in the clamping
area. Additionally, please follow the legal regulations valid in your country and the hygiene regulations for
your doctor’s office or hospital.
•
The SonoTT
TM
Clamp-On Transducer may not be submerged into cleaning or disinfecting
solution.
•
The SonoTT
TM
Clamp-On Transducer is not suited for cleaning processes using machines.
Sterilization processes, especially steam sterilization or autoclaving, may not be used.
•
The SonoTT
TM
Ultrasonic FlowComputer and the SonoTT
TM
Clamp-On Transducer are used
in non-sterile areas in operating rooms. Due to this, cleaning and disinfection after every
use as well as before the first use is essential.
The following surface disinfectant is recommended:
Name
Manufacturer
Contact
Bacillol® AF
Hartmann
www.hartmann.de
6.7 Changing the SonoTT
TM
Clamp-On Transducer
If
the
SonoTT
TM
Clamp-On
Transducer
was
exchanged
before
switching
on
the
SonoTT
TM
Ultrasonic FlowComputer, then the new calibration parameters will be shown in the Settings Menu
of the SonoTT
TM
Ultrasonic FlowComputer after the next start-up sequence.
If the SonoTT
TM
Clamp-On Transducer is exchanged during the operation of the SonoTT
TM
Ultrasonic
FlowComputer, the alarm is activated with an error message. Once the new SonoTT
TM
Clamp-On Transducer is
connected, the alarm is reset and the calibration data can be seen in the Settings Menu.
•
Make sure that the SonoTT
TM
Clamp-On Transducer is attached at a different place on the
tube each time to avoid deformation.
•
Ensure that the calibration table used is the one needed for your respective application.
SonoTT
TM
Clamp-On Transducer
Copyright em-tec GmbH
D162-700 SonoTT
TM
Clamp-On Transducer - Operating Instructions V3.0