While it is possible to carry out measurements outside of the specified flow range, em-tec GmbH cannot
guarantee that resulting flow values will be within the given accuracy range.
The parameters of the connected SonoTT
TM
Clamp-On Transducer and the selected calibration table are
displayed in the
Settings-Menu
of the SonoTT
TM
Ultrasonic FlowComputer as shown in the following example:
Calibration
Table
Transdcuer
Serial No.
Tube Size
Tube
material
Medium
Medium
Temperature
Maximum Measured Value
0
SN 12345
3/8" x 3/32"
PVC
Blood
(based on water)
37 °C (98.6 °F)
10000 ml/min
Pleasue ensure that the tube size and material used corresponds to that shown on the display of the SonoTT
TM
Ultrasonic FlowComputer.
6.5 Disassembly and Storage of the SonoTT
TM
Clamp-On Transducer
In order to not impair the measurement, it is best to disassemble the SonoTT
TM
Clamp-On
Transducer once the measurement process has been completed.
To disconnect the SonoTT
TM
Clamp-On Transducer, follow these steps:
1.
Open the spring-loaded lock and swing back the lid.
2.
Remove the tube from the channel.
3.
Close the lid and carefully remove the 15-pin, high density D-Sub socket from the screw connection at the
back of the SonoTT
TM
Ultrasonic FlowComputer.
Once you have disconnected the SonoTT
TM
Clamp-On Transducer, check for any residue of the medium or
any other kind of dirt or grime and clean the sensor if necessary. Store the sensor according to storage
specifications. Make sure to store it in a place where it cannot fall off or be dropped and ensure that the plug
is dry.
SonoTT
TM
Clamp-On Transducer
Page 16
D162-700 SonoTT
TM
Clamp-On Transducer - Operating Instructions V3.0