BioProTT
TM
Flow Measurement System
Version 1.00
page 2 of 75
Brief Operating Instructions - Quick Guide BioProTT
™
Flow Measurement System
Safety instructions
Chapter 2 - Page 6 ff.
Familiarize yourself with the safety instructions!
Setting up and connecting the flow meter
Chapter 6 - Page 15 ff.
Connect the BioProTT™ flow meter to power supply, e.g.
with supplied 24V DC power supply and connect analog or
digital interfaces as needed.
Installing flow sensor
Chapter 7 - Page 29 ff.
Connect Clamp-On Transducer to BioProTT™ flow meter
by plugging it into the correct socket.
Please ensure that transducer is proper size for the tubing.
Insert tubing into BioProTT™ Clamp-On Transducer and
close lid. The flow direction is indicated with an arrow.
Prepare Measurement
Chapter 7 - Page 29 ff.
Power the BioProTT™ flow meter on and wait approx. 15s
for initialization.
Usually a sensor will show a flow value even with no flow
present. Zero this (flow value) offset by pressing the
appropriate button or selecting the appropriate menu option.
Note: ensure that
correct calibration table is selected (for sensors with
more than one table programmed)
tubing is filled with medium,
there is sufficient coupling and
the liquid in the tubing not moving (no flow) before
adjusting the zero offset (flow value ml/min!)
Perform custom calibration if necessary
Perform Measurement
Chapter 7.7 - Page 32 ff.
The BioProTT™ flow measurement system is now ready
for use.
Please note: If troubleshooting is necessary please see chapter