BioProTT
TM
Flow Measurement System
Version 1.00
page 41 of 75
7.7.3
Measurements Using BioProTT™
Flow
Track plus
BioProTT™
Flow
Track plus is equipped with a display and supports the following
functions via the graphical user interface (GUI). The screens shown are only examples of
actual possible screen contents.
When using the display and changing settings like calibration table or
calibration factor in parallel via the digital interface the change via the
digital interface will only be visible at the display when the specific screen
is “flashed” i.e. if the user changes to a different screen and back. The new
value will however take immediate effect on the measurements.
7.7.3.1 Flow Value and Coupling
Flow value and coupling are displayed as follows.
Screen
Meaning
RSS 82%
Flow
320
Table 1
ml/min
Flow value and coupling are displayed as follows.
The flow value is calculated using the selected calibration table
and calibration factor.
When the flow value is greater than 1 l/min, the units displayed
are changed to [l/min].
In order to adjust the “zero” offset, please ensure that there is no
flow in the tubing and press the “zero” button until the flow
value is reset to zero.
7.7.3.2 Selection of Calibration Table
A Clamp-On Transducer can hold up to 7 calibration tables, which can be selected via the
appropriate screen, using the arrow and enter keys.
Screen
Meaning