BioProTT
TM
Flow Measurement System
Version 1.00
page 63 of 75
EMC:
Safety:
Conforms to:
EN 55011, CISPR 11 Class B, Group 1
CISPR 22 Class B,
47 CFR 15 / B,
IEC/EN 61326-1
Conforms to:
EN 60950 / IEC 60950
Power supply unit shipped with flow meter:
Type:
Friwo FW 7660, GPP 10, part number 1827497
Input voltage / current:
100-240 VAC (± 10%) / 50-60 Hz, max. 250 mA
Output voltage:
24 VDC (± 5%), 420 mA
Power consumption:
max. 10 VA
Fuse protection:
none
Power supply connection:
EMC:
Safety:
Worldwide use with exchangeable primary adapters
Conforms to:
EN 55011, EN 55022 / B,
FCC 47 part 15, EN 61000-3-2,
EN 61000-4-2, EN 61000-4-3,
EN 61000-4-4, EN 61000-4-5,
EN 61000-4-6, EN 61000-4-11
Fulfils Class II for the following applications:
EN 60950 / IEC 60950, UL 60950,
CSA 950 (cUL), VDE, CE label
11.2
Technical Data BioProTT™ Clamp-On Transducer
BioProTT
TM
Clamp-On Transducer:
Resolution
1ml/min
Accuracy*
± 3% of reading ± max. 20 ml/min
Measurement range
Dependent on sensor size and calibration (see
calibration certificate)
Dimensions (HxWxD)
Total weight (incl. cable/plug)
Weight sensor head
25 x 33 x 45 mm (ID ≠ 1/2”),
27 x 38 x 51 mm (ID = 1/2”)
52 g (ID ≠ 1/2”), 72g (ID = 1/2”)
52 g (ID ≠ 1/2”), 72g (ID = 1/2”)
Housing / lid material
Epoxy resin, aluminium, stainless steel
Cable length
2.9 m
Connecting plug
16-pin high density circular connector
IP code
IP65
Fluid types
Water, Buffer Solutions, Cell Culture Media,
Fermentation Media, Non-Aerated Liquids
Tubing type
e.g. Silicon, PVC and Polyurethanes
including Tygon®, C-flex®, Pharmed®, Cobe
®