EN - 51

liberoCONFIG
LI6003E

CONFIGURATION OF A LIBERO CX

3.8.2.3

Example of a configuration report

The following line numbers refer to the line of text in the "Placeholder Editor"

The used control characters quotation marks ("...") and semicolon (;) are interpreted 
according to the country settings of MS Excel. Depending on the country settings different 
characters for the described functions have to be used.

Notes

Placeholder Editor 
examples

Report opened with 
MS Excel

Report opened with 
an editor

Example of a report

Line Description

1

Date and time configured with the placeholder: Configuration time ("Format") con-
figured.

2

Text

3

-

Always use placeholder with quotation marks. In this case a semicolon (;) 
used within the text of the placeholder, will not be interpreted as column 
formatting.

-

Semicolon (;) between placeholders are used for column formatting.

-

Control characters have no effect on reports opened with an editor.

4

Entire line in quotes

-

Text and parameters are not separated into different columns.

-

Control characters have no effect on reports opened with an editor.

Summary of Contents for Libero Cx

Page 1: ...LIBERO C PDF Logger Indicator Operation Manual...

Page 2: ...al conditions All main items of the software product comply with the operating instructions and the information in the help file The guarantee does not apply to consumables disposable batteries or any...

Page 3: ...12 3 2 1 EoL End of Life 13 3 2 2 Schedule Start to stop of logging 14 3 2 3 Error codes 14 3 2 4 Temperature alarm 14 3 2 5 Configuration mode 15 3 3 Start of liberoCONFIG 16 3 3 1 Options 18 3 3 2 M...

Page 4: ...rt settings 41 3 8 1 Items 43 3 8 2 Configuration Report 46 3 8 3 Create SmartStart Pack Go 52 3 9 Help 53 4 Using SmartStart Pack Go 54 4 1 Change of Settings 56 4 2 Error Messages 57 5 Miscellaneous...

Page 5: ...nsport information is created as pdf file liberoCONFIG Program for the configuration of a LIBERO LIBERO SmartStart Application for simple and reliable assignment of config uration profiles on a LIBERO...

Page 6: ...LIBERO SmartStart appli cation Transfer is absolutely reliable and safe as no device settings are accessible during this process LIBERO SmartStart is used to load the correct configuration profile on...

Page 7: ...ned Up to three low alarm zones L can be defined PDF report with graph LIBERO CB PDF logger Single alarm time and MKT alarm Only one high alarm zone H and one low alarm zone L can be defined PDF repor...

Page 8: ...e from ELPRO BUCHS AG Lithium manganese dioxide Non rechargeable button cell battery No shipping declaration necessary 1 3 Disposal Europe WEEE U S A Find a drop off center for electronic waste in you...

Page 9: ...ations are on file at ELPRO BUCHS AG 3 4 5 1 Status information and error messages in the PDF report CE EMI LIBERO Cx PDF loggers and indicators meet the requirements regarding electromagnetic interfe...

Page 10: ...d value corresponds with the Profile ID see 3 4 6 Handling Options Profile ID 4 Alarm indicator The alarm indicator shows whether everything was within limits to date or if limits were violated 5 Temp...

Page 11: ...art Start Pack Go to quickly and safely assign configuration profiles to a large number of LIBEROs 3 1 1 Configuration file till SmartStart Pack Go exe file The configuration of an individual LIBERO C...

Page 12: ...CX 3 3 1 2 System requirements Windows XP 7 or 8 CPU 1 5GHz Memory 512 MB RAM Free hard disk space 100 MB Monitor 800 x 600 Pixel 3 2 Status information LIBERO Cx display LIBERO Cx Display Information...

Page 13: ...the PDF report but can be evaluated with elproVIEWER If the end of transport is marked by connection to a USB port no more data is logged Depiction of an alarm or error message Alarm indicator 3 3 St...

Page 14: ...er Ser vice 3 4 5 1 Status information and error messages in the PDF report Exxx Logging status is not changed PDF report evaluation is possible Displayed error code remains visible E210 Configuration...

Page 15: ...play Configured LIBERO Cx Either Strt or the Profile ID is shown on a configured but not started LIBERO Cx Start configuration As soon as the LIBERO Cx is connected to the USB port it switches to conf...

Page 16: ...eroCONFIG with 3 LIBERO Cx The following infor mation is shown LIBERO ID Type Firmware Version Expiry date Single multiple device selection Except Edit configuration all functions described in the fol...

Page 17: ...locations Password length Specify the paper format for the reports 3 3 1 Options Shows the information about the current program version and license number Refresh Refreshes the start window of liber...

Page 18: ...ngs Language German English Spanish French Italian Dutch Default File Loca tions 3 different file locations may be defined or selected For the configuration profiles see 3 4 Configuring the LIBERO Cx...

Page 19: ...tions Additional Settings Parameters Data entry field used for the definition of the minimum password length View Print Template for Calibra tion Data Paper Size You may choose between A4 and Letter N...

Page 20: ...ting up LIBERO Cx Save Profile Used to save a new profile in a file Apply The currently displayed configuration settings are transferred to the LIBERO Cx 3 3 4 Confirmation All successful actions are...

Page 21: ...ove the validity of the profile and as reference value for a con figuration check It is displayed in the Configuration window and is documented in the sec tion LIBERO Configuration of the PDF report 3...

Page 22: ...text fields limited to 15 characters fields 1 3 5 and 7 2 4 text fields limited to 25 characters fields 2 4 6 and 8 3 6 lines of 80 characters Hidden Information Two lines with text limited to 80 cha...

Page 23: ...starting the LIBERO Cx by pressing the Start Stop button data logging only starts when the delay time has elapsed dELY is shown on the display before logging is started and Run flashes Logging can al...

Page 24: ...erent alarm modes Single alarm zone Multi alarm zone The available alarm mode depends on the employed type of the LIBERO Cx A checked box activates the alarm mode 3 4 3 1 Single alarm zone LIBERO CB O...

Page 25: ...es T C or T F Data entry fields for the threshold values Alarm after An alarm is not triggered until the threshold violation has lasted longer than the speci fied time Event Single The delay time rest...

Page 26: ...one Tempera ture range C Alarm duration Add up time h Number of thresh old violations Plot section used for calculation H4 over 45 2 5 1 K H3 over 30 5 5 3 J K L H2 over 15 10 5 6 I J K L M W H1 over...

Page 27: ...value 83 kJ mol MKT Alarm Enable MKT alarm A checked box activates the alarm mode Lower MKT Limit Data entry field for the threshold value Upper MKT Limit Data entry field for the threshold value Onl...

Page 28: ...ration Alarm A checked box activates the alarm mode An alarm is triggered if the logging time oversteps this value Maximum Duration Data entry field for the threshold value Maximum possible alarm dura...

Page 29: ...port This setting is based on UTC Daylight Saving Time A checked box activates switchover between daylight saving time and winter time Data input fields for Start date End date Offset difference betwe...

Page 30: ...caled according to the lower and upper threshold values PDF Report Custom izing The following customizing options are activated by a checked box Hide alarm conditions in the PDF report Hide the alarm...

Page 31: ...t creation 4 Time of report creation 5 Device ID Logging Results File created This timestamp corresponds to the time at which the LIBERO Cx was connected to a USB interface for the first time WARNING...

Page 32: ...O CX 3 3 4 5 2 Example PDF report Hide alarm conditions in the PDF report Hide logging results in the PDF report Hide the alarm status in the PDF report Hide the chart in the PDF report Only LIBERO CS...

Page 33: ...g duration Enter the minimum logging time During this time the LIBERO Cx cannot be stopped with the Start Stop button or by connecting it to a USB interface Disable stop button When the box is checked...

Page 34: ...d by briefly pressing the Start Stop button less than 1 second Always show Profile ID instead of Strt Never show the Profile ID Default setting C001 3 4 7 Apply Configuration Profile All selected LIBE...

Page 35: ...EN 35 liberoCONFIG LI6003E CONFIGURATION OF A LIBERO CX 3...

Page 36: ...assword is set the data of the PDF report can only be evaluated if the password is known Set Change Reset If New Password Retype new Pass word are left blank the password will be reset If you have for...

Page 37: ...G LI6003E CONFIGURATION OF A LIBERO CX 3 3 6 Calibration View Print Creates a validation certificate for all selected LIBERO Cx This contains as many pages as LIBERO Cx selected Example Validation cer...

Page 38: ...38 EN liberoCONFIG LI6003E CONFIGURATION OF A LIBERO CX 3...

Page 39: ...the PDF report If the files have passed test results will be shown and can be printed or archived Procedure LIBERO PDF Report Integrity 1 Select and open PDF files which should be checked 2 Check PDF...

Page 40: ...40 EN liberoCONFIG LI6003E CONFIGURATION OF A LIBERO CX 3...

Page 41: ...lds to be defined so that ship ment specific information can be easily added i e carrier tracking number and viewable on the PDF report During the final programming process information can be added ei...

Page 42: ...liberoCONFIG LI6003E CONFIGURATION OF A LIBERO CX 3 Start LIBERO SmartStart Settings 3 8 1 Items 3 8 2 Configuration Report 3 8 3 Create SmartStart Pack Go File Return to liberoCONFIG File format libe...

Page 43: ...EN 43 liberoCONFIG LI6003E CONFIGURATION OF A LIBERO CX 3 3 8 1 Items Manage items Overview LIBERO SmartStart Settings...

Page 44: ...tle The title is the designation for the selected item Enter a short clear name Color Each item can be assigned a color The color serves to identify the item quickly at a later point while working wit...

Page 45: ...ormation of the PDF report during profile assignment no access There are no entries possible If all entries are set to no access no entry prompts appear during profile assignment with SmartStart Pack...

Page 46: ...gurations done by LIBERO SmartStart With the aid of the Placeholder Editor the used file name path and the parameters documented may be customized Reporting ON OFF Configuration report ON The logging...

Page 47: ...ettings Path Default setting for saving the configuration report Subdirectories Filename This button is used to insert placeholders for the designation of subdirectories and file names Report Content...

Page 48: ...elected placeholders determine the con tent of the PDF report 3 8 2 2 Placeholder Editor Example Placeholders are expressions that appear as variable information in the report Data logger ID This is t...

Page 49: ...nfiguration Format settings according to regional settings of the computer Time zone corre sponds to PDF settings Configuration Time Format Configuration Time Format Time of the configuration of the L...

Page 50: ...on Name ApplicationName Corresponds to the Window Title 3 8 3 Create SmartStart Pack Go Place holder Function dd Day with leading zero MM Month with leading zero MMM Shortcut month Jan Feb Mar MMMM Na...

Page 51: ...les Report opened with MS Excel Report opened with an editor Example of a report Line Description 1 Date and time configured with the placeholder Configuration time Format con figured 2 Text 3 Always...

Page 52: ...the configuration of a LIBERO Cx with a given profile without any addi tional programs and without installation File format exe Pack Go runs on any PC from a WEB FTP or file server or from the Intern...

Page 53: ...s and LIBERO SmartStart settings 3 4 Configuring the LIBERO Cx 3 5 Passwords 2 Create an exe file which includes the following information Defined profiles and LIBERO SmartStart settings File name and...

Page 54: ...ned in the LIBERO SmartStart settings 3 8 LIBERO SmartStart settings Mode The LIBERO Cx must be in Conf or Strt mode The first item from the LIBERO SmartStart list is displayed showing the assigned co...

Page 55: ...pen The infor mation can be entered via the keyboard or barcode reader Key to entries 1 Edit text 2 Add dropdown 3 Insert dropdown 4 Add text 3 8 1 Items Missing variable Error message When all entrie...

Page 56: ...B port and continue with the next LIBERO 4 1 Change of Settings A SmartStart Pack Go can contain several items profiles Click the setting window with the right mouse button and select the right profil...

Page 57: ...SING SMARTSTART PACK GO 4 4 2 Error Messages In case of an error a message will be displayed Reasons for error massages Profile incompatible Cancel selected during data download LIBERO has been discon...

Page 58: ...o com The license key required for licensing is sent to you free of charge e mail License Download The download link is also contained in the e mail The software can be downloaded up to twice with the...

Page 59: ...the following information Software version Select Info in the menu option Application Used LIBERO type PDF report of the LIBERO Which actions were carried out before the problem arose exact descripti...

Page 60: ...60 EN liberoCONFIG LI6003E MISCELLANEOUS 5...

Page 61: ...EN 61 liberoCONFIG LI6003E MISCELLANEOUS 5...

Page 62: ...EMI 9 EoL End of Life 13 Error message 14 Exceptional environmental conditions 9 Expiry date 10 16 53 F Factory settings 15 File format 21 42 52 File name 22 31 Format 19 Format Measurement Value 30 I...

Page 63: ...uration report 46 SmartStart items 44 SmartStart Pack Go 52 SmartStart password 44 SmartStart Placeholders 49 50 SmartStart report 51 Start 10 Start delay 23 24 Start options 14 Start Stop button 10 S...

Page 64: ...Operation Manual liberoCONFIG LI6003E 08 2013 ELPRO BUCHS AG Lang ulistrasse 62 CH 9470 Buchs SG Switzerland email swiss elpro com For local representatives see www elpro com...

Reviews: