EN - 25
liberoCONFIG
LI6003E
CONFIGURATION OF A LIBERO CX
3
3.4.3.2
Multi alarm zones,
LIBERO CS and CI
The function "Multi
alarm zones" splits the
alarm conditions in up
to 8 independent
zones; H1 - H4, G and
L1 - L3. Zone G repre-
sents the temperature
range in which none of
the alarm conditions
are fulfilled.
Temperature alarm activated for LIBERO CS and LIBERO CI with multi alarm zones
Alarm Zones
• Used
These check boxes are used to select the desired alarm zones.
• T[°C] or T[°F]
Data entry fields for the threshold values.
• Alarm after
An alarm is not triggered until the threshold violation has lasted longer than the speci-
fied time.
• Event
• Single
The delay time restarts for each threshold violation.
• Cumulative
An alarm will be triggered as soon as the elapsed time of all violations together has
reached the delay time.
• Excursions
Accepted number of excursions, without respect to the set alarm delay time.
• Zone H1 + L1 coupled.
If this field is checked, the information entered for the alarm zone in H1 are used for
both H1 overstepping and L1 understepping. (Total time and number of excursions
overstepping and understepping).