06-09-2016 10:40:26
2-2 Specifications (23±5
℃
)
A. pH / mV
pH electrode
Optional. Any PH electrode with BNC connector
Measurement
pH
0 to 14 pH
mV
-
1999 mV to 1999 mV
Input Impedance
10^12 ohm
Temperature
Compensation for
pH measurement
Manual
0 to 100
℃
, be adjusted by push button on front panel.
Automatic (ATC)
With the optional temperature probe ( TP-07 ) 0 to 65
0
C
pH Calibration
PH7, PH4, and PH10, 3 points calibration ensure the best linearity and accuracy.
Optional probe and
accessories.
* PH electrode:PE-03, PE-11, PE-01, PE06HD, PE-04HD, PE-05T, PE-03K7
* ATC (automatic temperature probe)
TP-07
* buffer solution:
pH 7
PH-07
pH 4
PH-04
ORP electrode
ORP-14
Measurement
Range
Resolution
Accuracy
pH
0 to 14 pH
0.01 pH
± (0.02 pH + 2 d)
mV
-1999 to 1999 mV
1 mV
± (0.5% + 2 d)
NB! PH accuracy is based on calibrated meter only.
B. Conductivity
Conductivity probe
Optional, Carbon rod electrode for long life.
Function
* Conductivity (uS, mS),
* TDS (Total Dissolved Solids, PPM)
* Temperature. (
0
C,
0
F)
Temperature
Compensation
Automatic from 0 to 60
℃
(32 - 140
℉
),with temperature compensation
factor variable between 0 to 5.0% per C.
Probe Operating Temp.
0 to 60
℃
.
Probe Dimension
Round, 22 mm Dia. x 120 mm length.
Optional probe and
accessories
* Conductivity probe
CDPB-03
* Pt Conductivity probe.
CDPB-04
* 1.413 mS Conductivity Standard Solution
CD-14
1. Conductivity (uS, mS)
Measurement
Range
Resolution
Accuracy
200 uS
0 to 200.0 uS
0.1 uS
± (2% F.S.+1d)
* F.S. - full scale
2 mS
0.2 to 2.000 mS
0.001 mS
20 mS
2 to 20.00 mS
0.01 mS
200 mS
20 to 200.0 mS
0.1 mS
* Temperature Compensation: Automatic from 0 to 60
℃
(32 - 140
℉
), with temperature compensation
factor variable between 0 to 5.0% per C.
* The accuracy is specified under measurement value
≦
100 mS.
* mS - milli Simens
* @ 23±5
℃