Elma 795
UK 40
v. 290817
06-09-2016 10:40:26
10. ADVANCEREDE INDSTILLINGER
Press and hold "
SET
" button (3-7, Fig.1) >2 sek. Subsequently press on "
SET
” button, once a while in
sequence to select the eight main function, the display will show
Adjust
Function
Default
value
Adjust and
save. Press
”Enter”
PH
OrP
do
O2
Cd
tdS
SALt
10-1
*Auto power OFF management
PoFF
PoFF
PoFF PoFF
yES
yES / no
10-2
*Beeper sound on/off
bEEP
bEEP
bEEP bEEP
yES
yES/ no
10-3
*Temperature unit
t-CF
t-CF
t-CF
C
0
C or
0
F
10-4
**DO salt% compensation
SALt
0
%
10-5
**DO height (meter) compensation
High-
0
Meter
10-6
**DO height (feet) compensation
HighF
0
Feet
10-7
**CD temperature compensation factor
PEr C
2.00
Grad
10-8
*pH temperature compensation value
t-SEt
25.0
Grad
Escape from the advanced setting.
Wait a few seconds or press "
ESC
”
(3-3, Fig.1), button and return to normal
screen.
ESC
ESC
ESC
ESC
NB!
Use the
"▲" (3-5, Fig. 1) and "▼" (3-6, Fig.1) to adjust of value
*
Save new defaults value with the
”
Enter
” button
(3-8, Fig. 1)
.
**
Temporary setting accepted and saved with the
”
Enter
” button
(3-8, Fig. 1)
.
11. POWER SUPPLY from 9V DC ADAPTER
The meter also can supply the power supply from the DC 9V Power Adapter (optional). Insert the plug of
Power Adapter into "DC 9V Power Adapter Input Socket" (3-17, Fig. 1). The meter will permanent power
ON when use the DC ADAPTER power supply (The power Button function is disable).
12. BATTERY REPLACEMENT
1.
When the left corner of LCD display show " ", it is necessary to replace the battery. However, in-
spec. measurement may still be made for several hours after low battery indicator appears before the
instrument become inaccurate.
2.
Loose the screws of the "Battery Cover" (3-11, Fig.1) take away the "Battery Cover" from the
instrument and remove the battery.
3.
Replace with DC 9 V (006 P) Alkaline/heavy duty, and reinstate the cover.
13. SYSTEM RESET
If the meter happen the troubles such as:
CPU system is hold (for example, the key button cannot be operated).
Then make the system RESET will fix the problem. The system RESET procedures will be
either following method:
During the power on, use a pin to press the "
Reset
Button" (3-18, Fig. 1) once a while will reset
the circuit system.