The consideration of Diaphragm (probe head with diaphragm set)

 

 

The oxygen probe component is the thin Teflon diaphragm housed in the tip of the probe. The 
diaphragm is permeable by the oxygen molecules but not by the considerably larger molecules 
contained in the electrolyte. Due to this characteristic, the oxygen may diffuse throughout the electrolyte 
solution contained in the probe, and the measurement circuit may quantify its concentration. 
 

 

 

 

 

 

 
1. Unscrew the "Probe head" (9-3, Fig 6). 
2. Pour out the old Electrolyte from the container of the "Probe head". 
3. Fill the new Electrolyte (OXEL-03) into the container of the "Probe head". 
4. Screw the "Probe head" (9-3, Fig 6) into the probe body. 

 

When not use the probe, should insert the " Probe head "into 
the " Probe protection cover "9-1 Probe handle( 8-5, Fig. 5 )

 

 

9-1 Probe handle 
9-2 Temp. sensing metal 
9-3 Probe head 

9. OTHER FUNCTION 

9-1 Data Hold 

During the measurement, press the "

Hold

 Button" (3-3, Fig.1) once will hold the measured value & the 

LCD will display a "

HOLD

" symbol. Press the "

Hold

 Button" once again will release the data hold 

function. 

 

9-2 Data Record (Max., Min. reading) 

 

1. The data record function records the maximum and minimum readings. Press the "

REC

 Button " (3-4, 

Fig.1) once to start the Data Record function and there will be a " 

REC

. " symbol on the display.. 

2. With the " 

REC

. " symbol on the display: 

a. 

Press the "

REC

 Button" (3-4, Fig.1) once, the "

REC

. MAX." symbol along with the maximum value 

will appear on the display. If intend to delete the maximum value, press the "Hold Button" (3-3, Fig.1) 
once, then the display will show the "REC." symbol only & execute the memory function continuously. 

b. 

Press the "REC Button"(3-4, Fig.1) again, the " 

REC

. MIN. " symbol along with the minimum value 

will appear on the display. If intend to delete the minimum value, press the "

Hold

 Button"( 3-3, Fig.1) 

once, then the display will show the "REC." symbol only & execute the memory function continuously.

 

c. 

To exit the memory record function, just press the "

REC

" button for 2 seconds at least. The display 

will revert to the current reading. 

 

9-3 LCD Backlight ON/OFF 

After power ON, the "LCD Backlight" will light automatically. During the measurement, press the 

"

Backlight

 Button" (3-2, Fig.1) once will turn OFF the "LCD Backlight". Press the "

Backlight

 Button" 

once again will turn ON the "LCD Backlight" again.  

 

Change the electrolyte 

Summary of Contents for ELMA 795

Page 1: ...Manual ELMA 795 Dansk Norsk Side 1 14 Svenska Sida 15 28 English Page 29 42 EAN 5706445230341...

Page 2: ...ORSYNING MED 9V DC ADAPTER 12 12 UDSKIFTNING AF BATTERI 12 13 SYSTEM RESET 12 14 RS232 PC SERIEL INTERFACE 13 15 VALGFRI TILBEH R 14 Elma 795 er et smart multi instrument som efter valg af probe kan m...

Page 3: ...e m ling Opl st ilt Brug den polargrafiske ilt probe med temperaturf ler til h j pr cisions m ling af opl st ilt DO og temperatur Kraftig probe til opl st ilt med probe hoved der kan monteres i BOD fl...

Page 4: ...RS 232 med kablet UPCB 02 ekstra tilbeh r Drift temperatur 0 til 50 Drift fugtighed Mindre end 85 R H Str mforsyning Alkaline DC 9 V 006 P batteri eller lignende DC 9V adapter indgang AC DC str mfors...

Page 5: ...0 5 2 d NB PH n jagtigheden er baseret p et kalibreret instrument B Ledningsevne Ledningsevne probe Valgfri grafitstang elektrode med lang levetid Funktion Ledningsevne uS mS TDS Total Dissolved Solid...

Page 6: ...emperatur kompensering Automatisk fra 0 til 60 0C 32 140 0F med variabel temperatur kompenseringsfaktor mellem 0 til 5 0 pr 0C Probe drifts temperatur 0 til 60 0C Probe dimensioner Rund diameter 22 mm...

Page 7: ...ik 3 16 PH stik BNC stik H jre side 3 17 DC 9V stik til str mforsynings adapter 3 18 Reset knap 3 19 RS 232 USB udgangs terminal 4 FUNKTIONER 1 T nd instrumentet ved et tryk p POWER tasten 3 2 Fig 1 a...

Page 8: ...derste display viser temperaturen m lt med ATC proben TP 07 NB N r pH elektroden ikke er i brug skal probehovedet 5 3 Fig 2 i opbevarings flasken 5 4 Fig 2 5 3 mV M ling Instrumentet har en indbygget...

Page 9: ...ibreringsproceduren som beskrevet i pkt 6 3 6 1 Ledningsevne m ling 1 Tilslut ledningsevne proben CDPB 03 6 1 Fig 3 til CD stikket 3 15 Fig 1 2 Brug Function tasten step til Cd funktionen Se pkt 4 3 H...

Page 10: ...indenfor 3 sek p Enter tasten 3 8 Fig 1 m lev rdien vises nu p b de verste og nederste display Brug 3 5 Fig 1 3 6 Fig 1 for at indstille v rdien svarende til kalibrerings v sken Tryk Enter 3 8 Fig 1 f...

Page 11: ...pkt 4 Displayet viser O2 og v rdien for ilt i luften der bruges som reference 8 2 Kalibrering 1 Tilslut proben 8 1 Fig 5 til DO stikket 3 14 Fig 1 2 Brug Function tasten step til O2 funktionen Se pkt...

Page 12: ...rdier 1 Gem funktionen gemmer m lte maksimum og minimum v rdier Tryk en enkelt gang p REC tasten 3 4 Fig 1 for at aktivere denne funktion REC symbolet vises p displayet 2 Med REC symbolet vist p displ...

Page 13: ...B Brug 3 5 Fig 1 og 3 6 Fig 1 for at indstille til nye v rdier Gem ny standardv rdi Indstilling godkendes og gemmes med Enter tasten Enter 3 8 Fig 1 Midlertidig indstilling godkendes og gemmes med Ent...

Page 14: ...13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0 Hver cifre indikere f lgende D15 Start bit D14 4 D13 N r data fra verste display sendes 1 N r data fra nederste display sendes 2 D12 D11 Annunciator for di...

Page 15: ...ik Str mforsyning AC 220V 230V til DC 9V Tysk stik PH valgfrit tilbeh r PH Elektrode 1 til 13 pH Model PE 11 PH Elektrode 1 til 13 pH Model PE 03 PH Elektrode 0 til 14 pH Model PE 01 Temperatur probe...

Page 16: ...LLNING 26 14 RS232 PC SERIELLT INTERFACE 27 15 VALFRIA TILLBEH R 28 Elma 795 r ett smart multi instrument som efter val av probe kan m ta flera olika kemiska enheter som tex surhet med pH och ORP v r...

Page 17: ...id m tning av konduktivitet Uppl st syre Anv nd den polargrafiska syreproben med temperaturgivare f r h g precisionsm tning av uppl st syre DO och temperatur Kraftig probe f r uppl st syre med probhuv...

Page 18: ...232 med kabeln UPCB 02 extra tillbeh r Drifttemperatur 0 till 50 Drift fuktighet Mindre n 85 R H Str mf rs rjning Alkaline DC 9 V 006 P batteri eller liknande DC 9V adaptering ng AC DC str mf rs rjni...

Page 19: ...ll 14 pH 0 01 pH 0 02 pH 2 d mV 1999 till 1999 mV 1 mV 0 5 2 d B Konduktivitet Konduktivitet prob Valfri grafitst ngselektrod med l ng livsl ngd Funktion Konduktivitet uS mS TDS Total Dissolved Solids...

Page 20: ...ng Temperaturkompensering Automatisk fr n 0 till 60 0C 32 140 0F med variabel temperaturkompenseringsfaktor mellan 0 till 5 0 per 0C Prob drifttemperatur 0 till 60 0C Prob dimensioner Rund diameter 22...

Page 21: ...O kontakt 3 15 CD kontakt 3 16 PH kontakt BNC kontakt H ger sida 3 17 DC 9V uttag f r str mf rs rjning Adapter 3 18 Reset knapp 3 19 RS 232 USB anslutning 4 FUNKTIONER 1 Sl p instrumentet genom ett tr...

Page 22: ...playen visar pH v rdet nedre visar temperaturen m tt med ATC proben TP 07 NB N r pH elektroden inte anv skall probhuvudet 5 3 Fig 2 ner i f rvaringsflaskan 5 4 Fig 2 5 3 mV m tning Instrumentet har en...

Page 23: ...alibreringsproceduren som beskrivs i pkt 6 3 6 1 Konduktivitet m tning 1 Anslut konduktivitet proben CDPB 03 6 1 Fig 3 till CD kontakten 3 15 Fig 1 2 Anv nd Function knappen stega till Cd funktionen S...

Page 24: ...m 3 sek p Enter knappen 3 8 Fig 1 m tv rdet vises nu p b de versta och nedersta displayen Anv 3 5 Fig 1 3 6 Fig 1 f r art st lla in v rdet svarande till kalibrerings v tskan Tryck Enter 3 8 Fig 1 f r...

Page 25: ...till O2 funktionen Se pkt 4 Displayen visar O2 och v rdet f r syre i luften som anv nds som referens 8 2 Kalibrering 1 Anslut proben 8 1 Fig 5 till DO kontakten 3 14 Fig 1 2 Anv nd Function knappen s...

Page 26: ...Spara data REC Max Min v rden 1 Spara funktionen sparar uppm tta max och min v rden Tryck en g ng p REC knappen 3 4 Fig 1 f r att aktivera denna funktion REC symbolen visas p displayen 2 Med REC symb...

Page 27: ...n NB Anv nd 3 5 Fig 1 och 3 6 Fig 1 f r att st lla in nya v rden Spara nytt standardv rde Inst llning godk nns och sparas med Enter knappen 3 8 Fig 1 Tillf lliga inst llningar godk nns och sparas med...

Page 28: ...3 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0 Varje siffra indikerar f ljande D15 Startbit D14 4 D13 N r data fr n versta displayen skickas 1 N r data fr n nedersta displayen skickas 2 D12 D11 Annunciat...

Page 29: ...20V 230V till DC 9V Tysk kontakt PH valfritt tillbeh r PH Elektrod 1 till 13 pH Modell PE 11 PH Elektrod 1 till 13 pH Modell PE 03 PH Elektrod 0 till 14 pH Modell PE 01 Temperaturprobe ATC prob Modell...

Page 30: ...UNCTION 39 10 ADVANCEREDE INDSTILLINGER 40 11 POWER SUPPLY from 9V DC ADAPTER 40 12 BATTERY REPLACEMENT 40 13 SYSTEM RESET 40 14 RS232 PC SERIAL INTERFACE 41 15 OPTIONAL ACCESSORIES 42 Elma 795 is a s...

Page 31: ...surement Dissolved oxygen meter use the polar graphic type oxygen probe with temperature sensor high precision measurement for Dissolved Oxygen DO and temperature measurement Heavy duty dissolved oxyg...

Page 32: ...optional accessories RS 232 cable UPCB 02 optional accessories Operating Temperature 0 to 50 Operating Humidity Less than 85 R H Power Supply Alkaline or heavy duty DC 9 V 006 P battery or equivalent...

Page 33: ...mV 0 5 2 d NB PH accuracy is based on calibrated meter only B Conductivity Conductivity probe Optional Carbon rod electrode for long life Function Conductivity uS mS TDS Total Dissolved Solids PPM Te...

Page 34: ...ure Compensation Automatic from 0 to 60 0C 32 140 0F with temperature compensation factor variable between 0 to 5 0 per 0C Probe Operating Temperature 0 to 60 0C Probe dimension Round Dia 22 mm x leng...

Page 35: ...ower Adapter Input Socket 3 18 Reset Button 3 19 RS 232 Output Terminal 4 FUNCTION SELECTION 1 Turn on the meter by pressing the Power Button 3 2 Fig 1 momentarily a Pressing the Power Button 3 2 Fig...

Page 36: ...show the setting manual Temp value NB When not use the Electrode it should immerse the Electrode sensing head 5 3 Fig 2 into the Protection bottle 5 4 Fig 2 5 3 mV Measurement The instrument build in...

Page 37: ...on procedures see pkt 6 3 6 1 Conductivity measurement 1 Connect the conductivity Probe CDPB 03 6 1 Fig 3 into the CD Socket 3 15 Fig 1 2 Use the Function Button step to the Cd function See pkt 4 3 Ho...

Page 38: ...g value will present on both upper and lower Display Use Button 3 5 Fig 1 Button 3 6 Fig 1 to adjust the up display value exact same as the standard conductivity value Press the Enter Button 3 8 Fig 1...

Page 39: ...ation 1 Connect the Probe optional DOPB 11 8 1 Fig 5 into the DO Socket 3 14 Fig 1 2 Use the Function Button step to the O2 function See pkt 4 3 Wait for 5 minutes at least until the display reading v...

Page 40: ...again will release the data hold function 9 2 Data Record Max Min reading 1 The data record function records the maximum and minimum readings Press the REC Button 3 4 Fig 1 once to start the Data Rec...

Page 41: ...emporary setting accepted and saved with the Enter button 3 8 Fig 1 11 POWER SUPPLY from 9V DC ADAPTER The meter also can supply the power supply from the DC 9V Power Adapter optional Insert the plug...

Page 42: ...splayed in the following format D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0 Each digit indicates the following status D15 Start bit D14 4 D13 When send the upper display data 1 When send the...

Page 43: ...lligent applications Power adapter AC 110V to DC 9V USA plug Power adapter AC 220V 230V to DC 9V Germany plug PH optional accessories PH Electrode 1 to 13 pH Model PE 11 PH Electrode 1 to 13 pH Model...

Page 44: ...Elma 795 v 290817 06 09 2016 10 40 26...

Reviews: