ELIS PLZE
Ň
a. s.
Design, Assembly and Service Manual
Page 11 of 44
Ultrasonic flow meter SONOELIS SE404X,
SONOELIS SE406X
ELIS PLZE
Ň
a. s., Lucni 15, P. O. BOX 126, 304
26 Plzen, Czech Republic, Phone: +420/377517711, fax: +420/377517722
Es90201K/c
4.2. Flow meters: basic technical specifications
Rated pressure of the measured fluid (PN)
40 or 16
Temperature of the measured fluid
0 to +150 °C (split version)
0 to +90 °C (compact version)
Ambient temperature
+5 to +55 °C
Maximum ambient relative humidity
80 %
Storage temperature
-10 to +70 °C at the relative humidity up to 70 %
Protection class
- evaluation electronic unit, compact version
IP 67
- evaluation electronic unit, split version
IP 65
- ultrasonic sensors UC 3.0, UC 5.0
IP 67
Sensor installed in piping
Flanges 11,
Č
SN EN 1092-1
Connecting cables for sensors
Standard length 5 m, maximum length 100 m
Max. difference in cable lengths
0.1 m
Electronic unit, split version
- dimensions
- weight
- power supply
- stand-by power supply
- power requirement
- mains fuse
- protection against electric shock,
Č
SN 332000-4-41
length 230 mm, height 217 mm, width 85 mm
1.5 kg
100
250 V, 50/60 Hz
3 V, Li battery (Lifetime 5 years)
6 VA
T 250 mA, 250 V
automated disconnection from power supply in the TN-S network
Evaluation electronic unit, compact version
- dimensions
- weight
- power supply
- stand-by power supply
- power requirement
- mains fuse
- protection against electric shock,
Č
SN 332000-4-41
length 152 mm, height 135 mm, width 205 mm
2.5 kg
100
250 V , 50/60 Hz
3 V, Li battery (Lifetime 5 years)
6 VA
T 250 mA, 250 V
automated disconnection from power supply in the TN-S network
Fluid flow velocity
minimum 0.1 m/s
maximum 10 m/s
Display (configurations STANDARD and COMFORT)
2 x 16-digit alphanumerical LC display
Outputs (optoelectronically isolated)
impulse, 0.1 to 1,000 l/imp (impulse length 50 ms)
frequency, 0 to 1,000 Hz or 10 kHz
(corresponds to flow rate 0 to q
s
)
switching output 24 V AC/0,1 A
Optional equipment
communication line RS 485
isolated current output 0
÷
20 mA or 4
÷
20 mA (corresponding to the flow
rate range of 0 to q
s
)
mass flow rate measurement accessory
modification for extended range of fluid temperature measurements; from -
20°C to +180°C (split version)
pipe-end flanges, flange packing pieces, bolts and nuts
sensor's protection IP 68
two direction flow measurement and direction indication