Design, Assembly and Service Manual 

Page 25 of 44 

Ultrasonic flow meter SONOELIS SE404X,  

SONOELIS SE406X 

 

ELIS PLZE

Ň

 a. s., Lucni 15, P. O. BOX 126, 304

 

26 Plzen, Czech Republic, Phone: +420/377517711, fax: +420/377517722 

Es90201K/c

 

Should the operator forget the password, it is possible to use the manufacturer’s password supplied with the 
system (0200). This shall be done as follows: switch off the power, depress and hold push-button T4 and switch 
the power on again. 
The user password can be changed at any time in the parameter setting mode using the procedure described in 
section 7.3.2.5 below. 
 

7.3.2.2. Meter setting procedures 

The meter parameters that can be defined or re-defined in the parameter setting mode include: the language of 
the messages appearing on the display, units of the displayed quantities, the user password required for entry 
into the parameter setting mode, the initial measured quantity, specified values of some measured quantities 
(q

s

, impulse number – litres per imp., threshold/sensitivity level, and the maximum/limit values of fluid flow rate, 

volume and temperature), as well as the date, day of the week, time of the day, start of the measurement period 
and meter zero position (only with the technological meters). 
The procedures to be used in setting particular parameters are described below. Upon initialisation of a 
particular parameter setting mode, the name of the function block concerned will appear on the first line of the 
display in block letters, e.g. 

 
 

   

 

       

LANGUAGE

 

 

 
At the same time, the current parameter name or value will appear on the second line. If you wish to pass on to 
the next parameter, depress T2; by depressing push-button T3 you will return to the previous parameter. Any 
parameter changes are done using push-button T1, confirmation of the new value by push-button T4. The 
display will then read 

 

  

PARAMETER SET 

 
 

To leave the current parameter setting mode and proceed to another parameter block, depress push-button T2. 
If you wish to return to the previous block, depress T3. To leave the parameter setting mode completely (and 
enter the data display mode for the parameter just set), depress push-button T4. 
 

7.3.2.3. Language selection 

The operator can choose from any of the six languages available (see Fig. 7.3.4, page 32). The language setting 
mode will be initiated as soon as the system acknowledges the correct user password. The first line on the 
display will then read  

 LANGUAGE 

 
 

or a message to the same effect in the actually defined language. On delivery, the language selected will be 
Czech unless the customer has specified their required language in the product order. The second display line 
will identify one of the languages available (e.g. Czech). Depress repeatedly push-button T1 to select the desired 
language. Upon selecting the language, confirm the setting by depressing push-button T4. The message on the 
display will inform the operator of completion of the parameter setting in the newly selected language. 

7.3.2.4. Measuring unit selection 

In this parameter setting mode, the desired measuring unit can be associated with each measured physical 
quantity (see Fig. 7.3.5, page 33). Upon initiating this mode, the first line of the display will read 

 

    

         

UNITS  

 
while the name a physical quantity will appear on the second line. Depress repeatedly push-button T1 to select 
the desired quantity and confirm by depressing T4. The quantity name will then appear on the first line and the 
second line will display one of the measuring units available. Select the desired unit by T1 and confirm by T4. 
Depress push-button T3 to access another measured quantity or use T2 to proceed to another parameter to be 
set. 
 

Summary of Contents for SONOELIS SE404X

Page 1: ...Manual Page 1 of 44 Ultrasonic flow meter SONOELIS SE404X SONOELIS SE406X ELIS PLZE a s Lucni 15 P O BOX 126 304 26 Plzen Czech Republic Phone 420 377517711 fax 420 377517722 Es90201K c Ultrasonic Fl...

Page 2: ...CONFIGURATION 19 7 2 THE STANDARD CONFIGURATION 19 7 2 1 Display data 20 7 2 1 1 Meter status information 20 7 2 1 2 Display of measured data 20 7 2 2 Review of the measured quantities 20 7 2 3 Review...

Page 3: ...ublic Phone 420 377517711 fax 420 377517722 Es90201K c 7 3 2 9 Meter zero setting 27 7 3 2 10 End of parameter setting 28 7 4 AUTOMATED METER TEST 28 8 SERVICE ACTIVITIES 36 8 2 WARRANTY SERVICES 37 8...

Page 4: ...een two ultrasonic transducers The flight times are measured for both directions of the signal wave propagation upstream and downstream whereby any asymmetry in the transducer positions are effectivel...

Page 5: ...fluid flow stabilisation piping sections at the meter input and output and generally increases the measurement accuracy In the case of a split version of the meter with separate sensor and the associa...

Page 6: ...he preset aggregate fluid mass level indication of measured values exceeding the preset temperature level meter failure indication The active state of the switching output can be selected either close...

Page 7: ...206 220 237 16 40 L mm 360 360 360 360 360 PN Nominal dim DN 100 DN 125 DN 150 DN 200 DN 250 DN 300 16 P mm 220 250 285 340 405 460 16 S mm 256 285 314 361 419 471 40 P mm 235 270 300 375 450 515 40 S...

Page 8: ...figuration 3 2 2 The compact version In the compact version the evaluation electronic unit is mounted directly onto the meter sensor UC 3 0 or UC 5 0 see section 3 2 1 1 above instead of the sensor te...

Page 9: ...ensure that no unauthorised modification of the meter functions or readings may take place The official certification and meter protection consist of Stick on labels with official seals of the respons...

Page 10: ...3 5 5 0 9 0 14 20 qNEC m3 h 0 07 0 09 0 14 0 22 0 37 0 6 0 9 1 2 2 2 3 5 5 Flow meter model SE406X DN 32 40 50 65 80 100 125 150 200 250 300 qs m3 h 20 32 50 80 150 240 350 500 900 1400 2000 qp m3 h 1...

Page 11: ...17 mm width 85 mm 1 5 kg 100 250 V 50 60 Hz 3 V Li battery Lifetime 5 years 6 VA T 250 mA 250 V automated disconnection from power supply in the TN S network Evaluation electronic unit compact version...

Page 12: ...selected sensor size can be re set in the COMFORT configuration using the customer control push button panel The graph below shows the pressure loss values at various flow rates for particular sensor...

Page 13: ...he ultrasonic flow meter shall be applied installed in observance of certain rules concerning meter placement in the fluid piping so as to ensure that the measurement accuracy complies with the meter...

Page 14: ...the sensor at all times Flow sensor location at the bottom pocket of the fluid piping If a flow sensor is to be installed in a vertical section of the fluid piping the fluid flow direction shall alway...

Page 15: ...and air bubbles either coming into the fluid through leaks in the piping or originating by the cavitation process in the sensor or other piping components If cavitation is suspected to appear in the s...

Page 16: ...ble conductors shall be soldered and the soldered joint protected against environmental and mechanical stresses by a suitable installation box All cables shall be led outside the thermal insulation la...

Page 17: ...C 5 0 W6 24V AC 0 1A RS 485 SWITCHING OUTPUT TRANSDUCER 1 TRANSDUCER 2 TRANSDUCER 3 IMPULSE OUTPUT OUTPUT CURRENT TRANSDUCER 4 50 60Hz Line power supply 100V 250V ON ON 2 FREQUENCY OUTPUT OFF 1 0 4 S2...

Page 18: ...ted to a relay coil in series with an external AC power source and so provide for indication of the flow direction or other selected parameter status Thermometer Pt 100 if used shall be connected to t...

Page 19: ...dges in the flow channel The separate box with electronic circuits in the case of a split version of the meter shall be attached to a suitable vertical support plate by means of four bolts of diameter...

Page 20: ...f the electronic unit will usually be manifested by discontinued regular changes of the system status symbols on the display 7 2 1 2 Display of measured data Up to three measured quantities can be dis...

Page 21: ...3 Review of the measured quantity units Volume flow rate Mass flow rate Volume Mass m3 hour t hour 1 000 m3 1 000 t m3 min t min m3 t m3 s t s l kg l hour kg hour 1 000 bbl 1 000 tons l min kg min bb...

Page 22: ...s plant Push buttons T1 to T4 are provided with the following graphic symbols T1 T2 T3 T4 The push button control of the meter is shown in a schematic diagram in see Fig 7 3 1 page 31 The system can...

Page 23: ...ows the ratio in per cent of the measured volume flow rate to the specified maximum volume flow rate 7 3 1 3 Mass flow rate The mass flow rate can be measured and the measured data displayed only on c...

Page 24: ...ould not perform measurements due to a failure condition 7 3 1 18 Power failure period The total duration of the period s in hours minutes and seconds from the measurement start during which the meter...

Page 25: ...hanges are done using push button T1 confirmation of the new value by push button T4 The display will then read PARAMETER SET To leave the current parameter setting mode and proceed to another paramet...

Page 26: ...nd the associated measuring unit will then appear on the first line of the display and the second line will show the previously defined limit value with the exception of the date and time The limit va...

Page 27: ...urement period the date hour and minute of the same The system will respond with a confirming message Parameter Set 7 3 2 9 Meter zero setting Before leaving the manufacturing plant every flow meter i...

Page 28: ...ro flow rate Then switch off the power depress push button S1 resetting the aggregate volume and with S1 depressed switch on the power again Upon releasing S1 the display will read TEST SENSOR FULL De...

Page 29: ...ual number of the minute of the cleaning procedure is displayed Every fourth seconds one numeral is added the line will be filled up by 15 same numbers within 1 minute after elapsing this time the dis...

Page 30: ...f physically possible values the following message will appear on the display UTS RATE LIMITS ADJUSTMENT and the actual limits will automatically be re adjusted with respect to the measured value The...

Page 31: ...Design Assembly and Service Manual Page 31 of 44 Ultrasonic flow meter SONOELIS SE404X SONOELIS SE406X ELIS PLZE a s Lucni 15 P O BOX 126 304 26 Plzen Czech Republic Phone 420 377517711 fax 420 377517...

Page 32: ...Design Assembly and Service Manual Page 32 of 44 Ultrasonic flow meter SONOELIS SE404X SONOELIS SE406X ELIS PLZE a s Lucni 15 P O BOX 126 304 26 Plzen Czech Republic Phone 420 377517711 fax 420 377517...

Page 33: ...Design Assembly and Service Manual Page 33 of 44 Ultrasonic flow meter SONOELIS SE404X SONOELIS SE406X ELIS PLZE a s Lucni 15 P O BOX 126 304 26 Plzen Czech Republic Phone 420 377517711 fax 420 377517...

Page 34: ...Design Assembly and Service Manual Page 34 of 44 Ultrasonic flow meter SONOELIS SE404X SONOELIS SE406X ELIS PLZE a s Lucni 15 P O BOX 126 304 26 Plzen Czech Republic Phone 420 377517711 fax 420 377517...

Page 35: ...Design Assembly and Service Manual Page 35 of 44 Ultrasonic flow meter SONOELIS SE404X SONOELIS SE406X ELIS PLZE a s Lucni 15 P O BOX 126 304 26 Plzen Czech Republic Phone 420 377517711 fax 420 377517...

Page 36: ...mbly and Service Manual Page 36 of 44 Ultrasonic flow meter SONOELIS SE404X SONOELIS SE406X ELIS PLZE a s Lucni 15 P O BOX 126 304 26 Plzen Czech Republic Phone 420 377517711 fax 420 377517722 Es90201...

Page 37: ...mmercial meter the parameters of a repaired product shall be verified at a duly authorised metrological centre 8 2 Post warranty services The post warranty services are understood to include any repai...

Page 38: ...on and maintenance manuals and a statement on the product compliance with the respective standards 13 WARRANTY CONDITIONS Unless agreed otherwise between the manufacturer and the customer the warranty...

Page 39: ...0 120 0 6 125 175 0 7 150 250 0 8 200 450 0 9 250 700 1 0 300 1000 1 1 non standard X X SN EN 1092 1 1 ANSI B 16 5 2 BS 4504 3 JIS B2210 4 non standard X carbon steel stainless steel painted of polyur...

Page 40: ...on standard X X FLOW METER SETTINGS one directional volume 1 one directional mass 2 bi directional volume 3 bi directional mass 4 non standard X standard qNEC according to manual 1 standard 10 mm s 2...

Page 41: ...on exceeding switch off 0 9 mass level on exceeding switch on 1 0 mass level on exceeding switch off 1 1 temperature level on exceeding switch on 1 2 temperature level on exceeding switch off 1 3 non...

Page 42: ...8 ton min 0 9 ton s 1 0 lb h 1 1 lb min 1 2 lb s 1 3 qp 1 4 non standard X X mass is not displayes 1 t 2 t 10 3 3 kg 4 ton US ton 5 lb libra 6 non standard X C 1 F 2 m s 1 ft s 2 Czech 1 English 2 Ge...

Page 43: ...n in 5 points with calibration certificate 5 standard calibration in 9 points without calibration certificate 6 standard calibration in 9 points with calibration certificate 7 standard calibration for...

Page 44: ...SONOELIS SE406X ELIS PLZE a s Lucni 15 P O BOX 126 304 26 Plzen Czech Republic Phone 420 377517711 fax 420 377517722 Es90201K c Address of manufacturer ELIS PLZE a s Lu n 15 P O BOX 126 304 26 Plze C...

Reviews: