Should the operator forget the password, it is possible to use the manufacturer’s password supplied with the
system (0200). This shall be done as follows: switch off the power, depress and hold push-button T4 and switch
the power on again.
The user password can be changed at any time in the parameter setting mode using the procedure described in
section 7.3.2.5 below.
7.3.2.2. Meter setting procedures
The meter parameters that can be defined or re-defined in the parameter setting mode include: the language of
the messages appearing on the display, units of the displayed quantities, the user password required for entry
into the parameter setting mode, the initial measured quantity, specified values of some measured quantities
(q
s
, impulse number – litres per imp., threshold/sensitivity level, and the maximum/limit values of fluid flow rate,
volume and temperature), as well as the date, day of the week, time of the day, start of the measurement period
and meter zero position (only with the technological meters).
The procedures to be used in setting particular parameters are described below. Upon initialisation of a
particular parameter setting mode, the name of the function block concerned will appear on the first line of the
display in block letters, e.g.
LANGUAGE
At the same time, the current parameter name or value will appear on the second line. If you wish to pass on to
the next parameter, depress T2; by depressing push-button T3 you will return to the previous parameter. Any
parameter changes are done using push-button T1, confirmation of the new value by push-button T4. The
display will then read
PARAMETER SET
To leave the current parameter setting mode and proceed to another parameter block, depress push-button T2.
If you wish to return to the previous block, depress T3. To leave the parameter setting mode completely (and
enter the data display mode for the parameter just set), depress push-button T4.
7.3.2.3. Language selection
The operator can choose from any of the six languages available (see Fig. 7.3.4, page 32). The language setting
mode will be initiated as soon as the system acknowledges the correct user password. The first line on the
display will then read
LANGUAGE
or a message to the same effect in the actually defined language. On delivery, the language selected will be
Czech unless the customer has specified their required language in the product order. The second display line
will identify one of the languages available (e.g. Czech). Depress repeatedly push-button T1 to select the desired
language. Upon selecting the language, confirm the setting by depressing push-button T4. The message on the
display will inform the operator of completion of the parameter setting in the newly selected language.
7.3.2.4. Measuring unit selection
In this parameter setting mode, the desired measuring unit can be associated with each measured physical
quantity (see Fig. 7.3.5, page 33). Upon initiating this mode, the first line of the display will read
UNITS
while the name a physical quantity will appear on the second line. Depress repeatedly push-button T1 to select
the desired quantity and confirm by depressing T4. The quantity name will then appear on the first line and the
second line will display one of the measuring units available. Select the desired unit by T1 and confirm by T4.
Depress push-button T3 to access another measured quantity or use T2 to proceed to another parameter to be
set.